Which of the following neuroscience methods measures the difference between oxygenated and deoxygenated blood in the brain?

Article Menu

Open AccessReview

1

Department of Physics/Electro-Optics Engineering, Jerusalem College of Technology, 21 Havaad Haleumi St., Jerusalem 91160, Israel

2

Monash Newborn, Monash Children’s Hospital, Melbourne 3168, Australia

3

Department of Neonatology, Shaare Zedek Medical Center, Shmuel Bait St 12, Jerusalem 9103102, Israel

*

Author to whom correspondence should be addressed.

Received: 5 August 2020 / Revised: 24 August 2020 / Accepted: 25 August 2020 / Published: 27 August 2020

Abstract

Adequate oxygen delivery to a tissue depends on sufficient oxygen content in arterial blood and blood flow to the tissue. Oximetry is a technique for the assessment of blood oxygenation by measurements of light transmission through the blood, which is based on the different absorption spectra of oxygenated and deoxygenated hemoglobin. Oxygen saturation in arterial blood provides information on the adequacy of respiration and is routinely measured in clinical settings, utilizing pulse oximetry. Oxygen saturation, in venous blood [SvO2] and in the entire blood in a tissue [StO2], is related to the blood supply to the tissue, and several oximetric techniques have been developed for their assessment. SvO2 can be measured non-invasively in the fingers, making use of modified pulse oximetry, and in the retina, using the modified Beer–Lambert Law. StO2 is measured in peripheral muscle and cerebral tissue by means of various modes of near infrared spectroscopy [NIRS], utilizing the relative transparency of infrared light in muscle and cerebral tissue. The primary problem of oximetry is the discrimination between absorption by hemoglobin and scattering by tissue elements in the attenuation measurement, and the various techniques developed for isolating the absorption effect are presented in the current review, with their limitations.

1. Introduction

Adequate oxygen delivery to tissue is essential for cellular metabolism, in which ATP energy-carrying molecules are produced from oxygen and nutrients. Since oxygen is transported from the lungs to the tissue cells via arterial blood, adequate oxygen supply to a tissue requires sufficient blood flow to the tissue and sufficient oxygen content in arterial blood, CaO2 [in mL oxygen per L blood]. The latter is related to the hemoglobin concentration in blood, [Hb] [g/L], and arterial oxygen saturation, SaO2, [defined as the ratio of oxygenated hemoglobin concentration to total hemoglobin in arterial blood] according to the following relationship [1,2]:

CaO2 [mL/L] = 1.36[Hb]SaO2 + 0.0031PaO2

[1]

where PaO2 is arterial oxygen tension [mmHg]. The factor 1.36 in Equation [1] stems from the maximal oxygen binding capacity of hemoglobin: 1.36 mL of oxygen in 1 g of hemoglobin. The value of 0.0031 PaO2 is the dissolved oxygen content in the arterial blood plasma, and is very small relative to the oxygen content carried by the hemoglobin.

Oxygen saturation, together with concentrations of oxygenated and deoxygenated hemoglobin [HbO2 and dHb, respectively], are quantitative parameters that allow the assessment of blood oxygenation. SaO2 provides information on the adequacy of respiration and ventilation. Oxygen saturation in venous blood, SvO2, also has clinical and physiological significance, because an increase in tissue metabolism requirements or a decrease in blood flow to the tissue lead to greater oxygen extraction by the tissue cells and consequently, to a decrease in oxygen saturation in the blood drained from the tissue by its veins. SvO2 is therefore directly related to the blood supply to the tissue that is drained by the vein and inversely related to its metabolism demand. The oxygen saturation in a tissue, StO2, is the ratio of oxygenated hemoglobin to total hemoglobin in the entire blood in the tissue—arterial and venous. StO2 is directly related to tissue blood flow and inversely related to its metabolism requirements. i.e., StO2 provides information on the adequacy of the blood supply to the tissue.

Oximetry is a technique for the assessment of blood oxygenation by optical measurements of light transmission through the blood, which is based on the different absorption spectra of HbO2 and dHb. Figure 1 presents the molar extinction coefficients of HbO2 and dHb [the latter is denoted as Hb in the figure] as a function of the wavelength, in the visible and near infrared wavelength regions [3] [the molar extinction coefficient is defined as the absorption constant per molar and is actually a specific absorption constant]. A great difference between the extinction coefficients of HbO2 and dHb appears for some wavelengths in the visible region, motivating their use for the oximetric assessment of blood oxygenation. However, light in the visible wavelength region below 650 nm is subjected to substantial absorption, and the transmitted light through tissue might have a low signal-to-noise ratio, even for a tissue with a small optical pathlength, such as the fingertip. Commonly, light in the red and infrared regions is utilized for measurements of light absorption in the fingertip and light in the near infrared region is preferred for measurements of light absorption in thicker tissue. Light in the visible region can also be used for oximetric measurements [4,5,6,7], mainly in the reflection mode.

The assessment of the oxygen saturation and HbO2 and dHb concentrations in the arterial and/or venous blood from measurements of light absorption [based on the different absorption spectra of HbO2 and dHb] is interfered by the scattering of the light when propagating through the tissue. The optical parameter measured in oximetry is attenuation, the reduction in intensity of light passing through the tissue, and both absorption and scattering contribute to the attenuation. Light is scattered when it is refracted at the border between two media with different refractive indices, such as the surface between intra and extracellular fluids. Scattering contributes to attenuation by the loss of light deflected from its trajectory, and also indirectly, by an increase in the optical pathlength and consequently, an increase in the absorption by the HbO2 and dHb molecules. The differentiation between the absorption and scattering components of the attenuation measurement is the primary problem in oximetry and various methods have been developed in the various oximetric techniques for isolation of the absorption component. The scattering constant in tissue and in whole blood monotonically decreases with light wavelength, but the slope is moderate relative to that of the hemoglobin absorption constant. Several studies present curves of reduced scattering coefficient µs’ in tissue [8,9] and in whole blood [10] as a function of light wavelength. µs’ is defined by µs’ = µs[1−g] where µs is the scattering constant and g is the anisotropy [11,12]. The path that the photon propagates before being scattered is represented by 1/µs; the path the photon propagates before entirely losing its direction is represented by 1/µs’.

The current review presents the main oximetric techniques that have been developed for the evaluation of oxygenation parameters in arterial, venous and tissue blood. The oxygenation parameters are derived from optical attenuation measurements, and the review mainly concentrates on the different approaches for isolating the absorption component from the entire attenuation effect. The degree of differentiation between the absorption and scattering effects is a pivotal determinant for the accurate assessment of oxygen saturation in any oximetric method.

2. Oximetry—General Theory

Oximetry is a technique for the evaluation of blood oxygenation level in arteries, veins, or in entire blood in tissue, by optical transmission measurements, utilizing the different absorption spectra of HbO2 and dHb. The blood oxygenation parameters are oxygen saturation and HbO2 and dHb concentrations in the arterial and/or venous blood. In order to measure the absorption, light with wavelength spectra in the red and/or infrared regions is irradiated into a tissue and its transmission through the tissue is measured. During its passage through the tissue, the light undergoes absorption, mainly by the hemoglobin molecules in the red blood cells, and scattering by tissue elements and the red blood cells. The light absorption that depends on the oxygenation level is the signal and the light scattering that is insensitive to the oxygenation level [10] is a noise, and suitable isolation of the contribution of the absorption to the light attenuation is essential for the accurate evaluation of blood oxygenation level. In order to derive the blood oxygenation from the light absorption measurement, the values of εO and εD, the extinction coefficients of HbO2 and dHb, respectively [Figure 1], are required, and they were obtained by measuring the absorption constant in scattering-free hemoglobin molecule solution, after fragmentation of the red blood cells [hemolysis] [13]. The transmitted light intensity, It, through a sample of homogeneous solution of hemoglobin molecules, is given by the Beer–Lambert law,

where I0 is the intensity of the incident light, d is the width of the hemoglobin solution [the source-detector distance], ε is the hemoglobin solution extinction coefficient and C is the hemoglobin concentration.

The transmitted light intensity, It, through a scattering tissue sample, is given by the modified Beer–Lambert law [12,14]:

It = I0exp[−G − εCl]; ln[I0/It] = G + εCl

[3]

where I0 is the intensity of the incident light and l is the mean optical path-length. l is greater than d because of scattering by the red blood cells and the tissue elements. The ratio l/d is named the differential pathlength factor [DPF] and it depends on the light scattering and absorption [12,14]. G represents the light-loss due to tissue scattering and depends on the light-sources and detector geometry. In the transmission mode of measurement, G represents the light-loss due to light deflection from a straight-line trajectory, while in the reflection mode [where the light-source and the detector are positioned on the same side of the tissue sample], G is more complex, as scattering also enables the diversion of light towards the detector.

The weighted mean extinction coefficient ε of the blood in the vessels of the tissue sample depends on the oxygen saturation SO2 in the tissue blood:

ε = εOSO2 + εD[1 − SO2].

[4]

Combining Equations [3] and [4], a linear relationship between the physiological parameter SO2 and the measured parameter ln[I0/It] can be obtained:

ln[I0/It] = G + εCl = G + εDCl + Cl[εO − εD] SO2.

[5]

C, [in units of molar], is either the hemoglobin concentration in tissue, when the measurement is done on blood in tissue, or the hemoglobin concentration in a blood sample in a blood vessel. If light is transmitted through a tissue, other substances, such as melanin, myoglobin and cytochrome, might also affect the absorption.

In order to derive SO2 from measurement of ln[I0/It] through Equation [5], the three unknowns, G, C and l, should be considered. The two effects of scattering, the light deflection and the lengthening of the optical path-length, considerably affect the light attenuation and prevent quantitative derivation of oxygen saturation from light transmission measurements. In addition, G and DPF change considerably among different persons, tissues and situations, and quantitative measurement of them in each examination is impractical. In oximetry, transmission measurement, in two or three wavelengths, has been utilized to reduce the harmful effect of the scattering in oximetry. The benefit of two-wavelength oximetry arises from the considerable spectral dependence of the difference in extinction coefficient between HbO2 and Hb, while the spectral changes in scattering are moderate [8,9,10].

3. Non-Pulsatile Oximetric Techniques

3.1. In-Vitro Measurements

Early in vitro oximetric measurements of oxygen saturation in whole blood [blood that contained intact red blood cells] were based on the assumption that light absorption and scattering can be treated independently—the theory of Twersky [15,16,17]. Then, the modified Beer–Lambert law can be written as OD = G + εcd, where OD is the optical density [OD = log10[I0/It] = ln[I0/It]/2.3]. In this equation, the source-detector distance, d, was used as the pathlength [instead of l in Equation [3]], neglecting the effect of scattering by red blood cells on pathlength elongation.

Pittman [16] isolated the contribution of scattering to attenuation by measuring the optical density at two wavelengths [546 and 520 nm], for which the extinction coefficients of dHb and HbO2 were equal [isosbestic wavelengths] and their scattering effect was similar, because they were close to each other. SO2 could be obtained from an additional measurement of the optical density at a third wavelength [555 nm] in which the extinction coefficients of Hb and HbO2 showed a large difference. SO2 was determined from the corrected optical densities at the wavelengths 555 nm and one of the isosbestic wavelengths: [OD[555]-G]/[OD[520]-G]. The three-wavelength technique with two isosbestic wavelengths was also used later in retinal oximetry [18,19,20], dealt with in Section 4.3.

Iwasaki et al. [17] measured the oxygen saturation of whole blood collected from rabbits by means of a fiber-optic catheter inserted in the blood [see Figure 2]. They used two wavelengths, an isosbestic one of 805 nm, and another of 670 nm, for which the extinction coefficients of dHb and HbO2 showed a great difference. Based on the theory of Twersky, they employed an expression:

SO2 = [k1/CHb][OD670-OD805] + k2

[6]

where CHb is the total hemoglobin concentration and k1 and k2 are constants that depend on the catheter geometry and hemoglobin extinction coefficients. OD670 and OD805 are the measured optical densities at 670 nm and 805 nm, respectively.

Equation [6] was validated in-vitro on a blood sample, making use of SO2 measured by a commercial oximeter [Radiometer], which employed two wavelengths in the visible range, 505 nm and 600 nm. The correlation between the two oximetric techniques was 0.99 [p < 0.001] for a saturation range of 13–100%, with standard deviation of 1.1%.

3.2. Tissue Oximetry

An oximeter that utilized red and green light transmission through the outer ear was developed during World War II [21]. Red light is more sensitive to the hemoglobin oxygenation level than broadband green light [Figure 1]. The study included a series of examinations, each comprised an initial calibration by comparing in vivo measurement during pure oxygen breathing, and measurement on extracted arterial blood with 100% saturation. Then, in vivo measurements were performed, in which the oxygen level in the inspired air was lowered. The calibration enabled the elimination of the scattering effects and the effect of light absorption by venous blood. The accuracy of the device, as determined by gas analysis of arterial blood samples, was 3–8%.

An ear oximeter that did not require calibration in each examination, but only pre-calibration, was developed by Hewlett-Packard in the 1970’s, [at about the same time as that of pulse oximetry invention], using eight wavelengths, from 650 nm to 1050 nm. The ear was chosen because its high blood flow is in excess of its metabolic needs, as it serves as a heat radiator. The coefficients required for the determination of arterial oxygen saturation from the optical measurements were derived by pre-calibration, using simultaneous measurements of light transmission and oxygen saturation in arterial blood extracted from volunteers that were performed in different oxygen saturation values, obtained by breathing different gas mixtures [22,23,24].

The calibration process could take account of the contribution of non-hemoglobin tissue absorbers and the scattering by the tissue and red blood cells. In order to compensate for changes in venous blood concentration, ear temperature was increased to 41 °C [or vasodilator cream was applied to the ear], “arterializing” the venous and capillary blood. The Hewlett-Packard eight-wavelength oximeter was compared to another oximeter [Radiometer], and showed a standard error of estimate of 2–2.5% [22,23,24].

The Hewlett-Packard oximeter has not been produced since 1983 [25], because the pulse oximeter that was invented in 1972 by Takuo Aoyagi [26] has been found to be more suitable for clinical practice. In a subsequent section, it will be described how pulse oximetry, after pre-calibration, solved the problems of interference of scattering and absorption in venous blood in a more practical way.

3.3. Intravenous Blood Oximetry

As described in the Introduction section, the oxygen saturation in a vein’s blood, SvO2, is related to the blood supply to the tissue that is drained by the vein. Of particular significance is the oxygen saturation in the pulmonary artery that carries the mixed venous blood—the entire venous blood returning from the body organs supplied by the systemic circulation—and in the jugular veins, which drain the blood from the brain. SvO2 in the pulmonary artery and in the jugular vein are directly related to the cardiac output and brain blood supply, respectively, and inversely related to the oxygen demand in the respective tissue.

Mixed venous oxygen saturation in the pulmonary artery can be obtained by introducing a Swan–Ganz catheter into the pulmonary artery and measuring the oxygen saturation in an extracted sample of mixed venous blood via blood gases analysis. Pulmonary artery catheters are introduced through a central vein, the right atrium and ventricle, and because their use has been limited, measurement of oxygen saturation in central veins—superior and inferior vena cava—has been suggested as a surrogate for mixed venous oxygen saturation measurement [27,28,29].

Monitoring oxygen saturation by intermittent measurement in extracted blood samples has several drawbacks, and intravenous blood oximetry has been developed for continuous monitoring of oxygen saturation in veins’ blood. The oximetric probe consisted of two optical fibers with a small gap between their tips [Figure 2], which was introduced into the vein. One of the fibers transmits two-wavelengths of light into the blood in the gap between the optical fibers and the second fiber guides the transmitted light to the detector [17,29,30].

The SvO2 value was calculated from the values of the transmitted light in two or three wavelengths by using the modified Beer–Lambert law [17,30]. Like the Millikan’s oximeter, the intravenous oximeter needs calibration in a blood sample before each continuous examination, according to the manufacturer’s instructions, using blood gases analysis. Limits of agreement of 7–16% were found between the optical measurement and blood gases analysis on blood samples extracted from the same vein [28,29].

Besides its use in the central veins, the intravenous blood oximeter has also been implemented for the measurement of oxygen saturation in the jugular vein, which drains blood from the brain. However, while low SvO2 in the jugular vein’s blood faithfully indicates inadequate oxygen supply to the brain, normal values of jugular oxygen saturation cannot ascertain adequate oxygen delivery to each brain site, because insufficient oxygen supply to a small region of the brain might only slightly affect SvO2 in the jugular vein, which drains blood from the entire brain hemisphere [31,32]. Several studies showed an association between abnormalities in jugular SvO2 and poor outcomes in patients after traumatic brain injury [33,34,35].

4. Pulse Oximetry—Theory and Techniques

4.1. Theory of Arterial Pulse Oximetry

The pulse oximeter, which measures the oxygen saturation in the arterial blood, SaO2, was invented in 1972 by Takuo Aoyagi, an electrical engineer at the Nihon Kohden company in Tokyo [26]. Pulse oximetry evaluates the SaO2 level by optical transmission measurements in two wavelengths, utilizing the different absorption spectra of HbO2 and dHb, like the other oximetric techniques. In order to measure oxygen saturation only in the arterial blood, the light absorption is measured selectively in the systolic increments of arterial blood volume after ventricular contraction, by recording heart-induced oscillations in light absorption—photoplethysmography [PPG].

The theory of pulse oximetry has been described in several publications [36,37,38,39,40]. In the current review, the theory and its assumptions will be presented in detail, in order to demonstrate its limitations and its relationship to the other oximetric techniques.

A finger PPG signal is shown in Figure 3, presenting the decrease in transmitted light intensity through the finger during systole and the light intensity increase during diastole, due to the systolic increase and diastolic decrease in the arterial blood volume. The maximal and minimal values of the PPG pulse [ID and IS in Figure 3, respectively] are proportional to the light transmission through the tissue at end-diastole and at end-systole, when tissue blood volume is minimal and maximal, respectively. The analysis of the PPG pulses enables the isolation of the arterial blood contribution to the entire light absorption.

Substitution of the ID and IS values for It in Equation [3], and division of the resultant two equations, eliminate the value of I0, and provide an equation for ID/IS:

IS = IDexp[GD + εCDlD]exp[−GS − εCSlS]; ln[ID/IS] = GD-GS + εCDlD − εCSlS

[7]

The D or S subscripts indicate the parameter values at diastole or systole, respectively. GD, GS, lD and lS are unknown, but it can be assumed that the changes in G and l due to the systolic arterial blood increment can be neglected relative to the absorption effect [GD ≈ GS, lD ≈ lS ≈ l]. Accordingly, Equation [7] can be written as:

The PPG amplitude, ID–IS, is generally small relative to IS, and Equation [8] can be written as:

where ΔC is the maximal hemoglobin concentration increase in the tissue during systole, due to the arterial blood volume increase. ε is the extinction coefficient for the arterial blood, which includes both oxi- and deoxi-hemoglobin, and depends on the oxygen saturation in the arterial blood, SaO2.

When light transmission is measured in two wavelengths, λ1 and λ2, Equation [9] becomes:

[[ID − IS]/IS]1 = ε1ΔC1l1; [[ID − IS]/IS]2 = ε2ΔC2l2

[10]

In pulse oximeters, the light in two wavelengths is emitted from two adjacent light-emitting diodes [LEDs] and the transmitted light is measured by a single detector. Since the light of the two wavelengths propagates through the same tissue element, the blood concentration change, ΔC, in the measurement site for the two wavelengths can be considered equal [ΔC1 ≈ ΔC2].

By dividing the equations in Equation [10] one gets:

R≡[[ID−IS]/IS]1 [[ID−IS]/IS]2=ε1 l1ε2l2

[11]

If we further assume that l1 is not much different than l2, then

Substitution of Equation [4] in Equation [12] provides a relationship between the measured parameter R and the physiological parameter SaO2:

R=εD1+SaO2[εO1−ε D1]εD1+SaO2[εO2−εD2 ]

[13]

and

SaO2 = εD1−Rε D2R[εO2−εD2]+[εD1 −εO1]

[14]

Equation [14] could provide the required relationship between SaO2 and R, since the values of εO and εD, the extinction coefficients for HbO2 and dHb, are known for wavelengths in the red and infrared regions [42,43]. However, the assumption that l1 is not much different than l2 [Equation [12]] can introduce significant error in the calculation of SaO2, when the two wavelengths differ significantly, since the optical path-length increases by light scattering and the scattering constant decreases monotonically with the wavelength [8,9,10]. In the commercial pulse oximeters that make use of two wavelengths in the red and infrared regions, a non-negligible difference in optical path-lengths between the two wavelengths is expected.

If the difference between l2 and l1 is not neglected, Equation [12] should be modified to

Furthermore, by replacing R in Equation [14] with R[l2/l1], the relationship between SaO2 and R becomes [39,41,44]:

SaO2 = εD1−R [l2/l1]εD2R[l2/l1][εO2 −εD2]+[εD1−εO1]

[16]

Generally, the factor l2/l1 is not known with reasonable accuracy and SaO2 cannot be derived from the measured value of R by Equation [16]. In practice, SaO2 is obtained from the measured parameter R by calibration, for each kind of pulse oximeter sensor [37,38,45,46]: R is measured for a number of subjects simultaneously with in vitro SaO2 measurement in extracted arterial blood by means of a co-oximeter [the gold-standard for SO2 measurements]. For each person, R and the corresponding SaO2 are measured for several values of oxygen partial pressure in the inspired air, and a look-up table is prepared from the R-SaO2 pairs. In clinical examinations, SaO2 is determined from the R value through an empirical function that was derived from the look-up table. The calibration enables the determination of SaO2 from the R measurement only, while the use of Equation [16] must utilize the value of l2/l1, which is not known with adequate precision.

Note that in general, in publications about pulse oximetry, the ratio of ratios, R, is defined by means of the pulsatile and non-pulsatile [average] components of the detected PPG pulse: R ≡ [AC/DC]1/[AC/DC]2, which seems to be different from the definition of R in Equation [11] [See Figure 3 for the definitions of AC and DC]. However, AC is equal to ID − IS and DC is the average value of the PPG pulse during the cardiac cycle. DC is only slightly greater than IS, assuming that ID − IS is small relative to IS.

The validity of the clinical measurements by the calibrated pulse oximeter is based on the assumption that the extinction coefficients and l2/l1 do not change significantly between different persons and different physiological and clinical situations. While the HbO2 and dHb extinction coefficients for a given wavelength are invariable, the scattering by tissue and red blood cells strongly depends on the substance and structure of the tissue and on the volume and anatomical arrangement of blood vessels in the tissue, which significantly varies among patients. It is expected that l2/l1 in a specific clinical examination might deviate from the mean l2/l1 value for the persons examined in the calibration process, and that deviation is probably the main origin of the inaccuracy in the assessment of SaO2 by pulse oximetry [See below]. For wavelengths in the red and infrared regions, the differences between their scattering constants and pathlengths are great [8,9,10] and accordingly, the absolute variation in the pathlengths ratio, l2/l1, among patients is expected to be considerable.

The actual mean error in SaO2 measurement by the available commercial pulse oximeters is 3–4% for adults and 4–6% for newborns [47,48,49,50,51,52,53,54]. In the low SaO2 levels, below 80 or 90%, the error is even greater [50,51,53]. Because of its low accuracy, arterial oxygen saturation measurement by pulse oximetry is denoted by SpO2, while the term SaO2 is generally reserved for arterial oxygen saturation measured by blood gases analysis in extracted blood. The accuracy in SaO2 measurement is of particular importance in preterm newborns treated with oxygen supplementation, because low oxygenation increases mortality while excessive oxygenation increases the risk of retinopathy of prematurity [53,55,56]. Accurate SaO2 measurement is also important in adult patients in intensive care units, treated with high concentrations of inspired oxygen, where hyperoxia is associated with greater mortality and morbidity [57,58].

In the last few years, pulse oximetry that makes use of two close wavelengths in the infrared has been developed [39,41,59]. The choice of light with two close wavelengths reduces the differences between their scattering constants and path-lengths, so that the error due to the inter-subject variation in l2/l1 is expected to be small relative to that experienced for red and infrared pulse oximetry. In those studies, SpO2 was derived from Equation [14], using the values of the extinction-coefficients of HbO2 and dHb, as presented in the articles of Kim and Liu [42,43], and neglecting the difference between l2 and l1. In the earlier study [39], simultaneous examinations of SaO2 were performed with two close wavelengths in the infrared region [767 and 811 nm] and two wavelengths in the red and infrared regions [635 and 937 nm], making use of Equation [14]. The discrepancy of the results from a commercial pulse oximeter [that was calibrated by extracted blood] was 3% for the two-infrared wavelengths pair and 9% for the red–infrared pair, probably due to the greater deviation of the ratio l2/l1 from the value of 1 in the red-infrared pair. The bottom line is that the two-infrared pulse oximeter also needs calibration, but since the difference between l2 and l1 is small, the variability in l2/l1 for the two-infrared pair of wavelengths is expected to be smaller than that for two wavelengths in the red and infrared regions.

In a recent study [41] on newborns and children with an arterial line, simultaneous measurements of SpO2, using two infrared wavelengths and SaO2 by blood gases analysis on extracted blood, were performed. The mean difference between SaO2 and SpO2 was 3%, which was corrected by multiplying the SpO2 results by a calibration factor of 1.03. Compared to arterial line measurements, the standard-deviation of the two-infrared-wavelength SpO2 measurements was 3.3%, similar to that of a commercial pulse oximeter [3.8%].

4.2. Pulse Oximetry of Venous Blood

Oxygen saturation in venous blood has been measured by modified pulse oximetry, in which a pulsatile increase in venous blood volume was achieved either by applying pressure proximal to the measurement site—e.g., a finger [39,60]—or by mechanical ventilation [61,62]. The latter technique is based on the phenomenon that mechanical ventilation induces systemic blood volume changes and on the assumption that those changes are of venous origin. SvO2 was measured by Walton et al. on cardiac surgery patients making use of an esophageal PPG probe, and by Wolf et al. on cerebral tissue of infants. The two studies yielded saturation values around 80% [62] or 70% [61], which are within the physiologic range of SvO2 in the relevant tissues.

In the two other studies [39,60], external pressure of 40–50 mmHg was applied proximal to the measurement site, leading to an increase in venous blood volume, and two-wavelength light absorption was used for the measurement of SvO2. Nitzan et al. [39] used a single step increase in the cuff pressure and two close wavelengths in the infrared region in order to reduce the difference in their pathlengths [see end of Section 4.1]. In the study of Khan et al. [60], artificial venous blood volume oscillations at a frequency of 0.2 Hz were induced by a pressure cuff, placed on the finger, proximal to a commercial pulse oximeter sensor. The interquartile range of the SvO2 difference between the venous pulse oximeter and a reference blood gas analyzer was −2.05–1.27%.

4.3. Difference Oximetry in Single Arteries and Veins

In pulse oximetry, the measurement of the attenuation difference between end-systole and end-diastole, given by ln[ID/IS], enables the determination of SaO2, by assuming that the difference in light-loss due to scattering between end-diastole and end-systole can be neglected relative to the absorption effect. In a similar way, difference oximetry was suggested for oxygen saturation measurements in a single blood vein, by introducing light in two wavelengths into the vein and measuring light reflection from two nearby sites in the vessel [63]. The method was found feasible by Monte Carlo simulation of light backscattering from a vein model.

In practice, difference oximetry has been implemented in oxygen saturation measurements in single retinal arteries and veins in the retina, by illuminating the eye fundus and acquiring its image through the eye’s pupil, mostly with a fundus camera, exploiting the eye’s transparency. Figure 4 presents light rays’ trajectories from the eye’s lens to the retina and scattered/reflected light from a retinal blood vessel and its neighborhood [See Hammer et al. [64]]. Most of the light that reaches the retina is lost by scattering or absorption in blood vessels, but a small part of the scattered/reflected light is recorded by the camera. Retinal image analysis enables discrimination between scattered/reflected light from a blood vessel [Iv] or from its neighborhood [I0n]. [Note that in retinal oximetry studies, scattered light from the vessel’s neighborhood is denoted by I0. We changed the symbol; as in the Beer–Lambert law, I0 denotes the incident light, see Equations [2], [3] and [5]] Retinal oximetry is based on the assumptions that I0n is unaffected by light absorption in the blood and the light-loss G due to scattering is similar for the two measurement sites, in analogy to pulse oximetry that assumes similarity in light-loss G for end-systole and end-diastole [see discussion after Equation [7]]. Hence, the difference between Iv and I0n is assumed to stem from the light absorption in the vessel’s blood, and the ratio between the optical densities for the two wavelengths enables the determination of oxygen saturation in the retinal vessel. It seems, however, that the assumption that I0n is unaffected by light absorption in the blood is not fully justified, as can be inferred from Figure 4.

The optical density of the vessel [ODVes] is defined as,

reflecting the effect of light absorption in the vessel’s blood. In practice, ODVes is obtained along the image of the vessel and next to it. The retinal image is split into two images created with two wavelengths in the visible region: λ1, in which the extinction coefficient substantially differs between HbO2 and dHb, and λ2 that is insensitive to oxygen saturation—an isosbestic wavelength. The image separation is performed either by an image splitter and appropriate filters [65,66], a double bandpass filter and the red and green camera channels [67,68], or a liquid crystal tunable filter [69]. The ratio between ODVes in the two wavelengths, ODVes1/ODVes2, is the measured parameter, from which the oxygen saturation is derived, assuming a linear relationship between the ODVes1/ODVes2 and SO2 [either SaO2 or SvO2],

SO2 = a + k[ODVes1/ODVes2].

[18]

a and k are constants that were determined by calibration, in which SO2 was measured in systemic arteries by a pulse oximeter, simultaneously with the retinal arteries examination, during varied hypoxia levels, obtained by breathing different oxygen concentrations [65]. In other studies [66,67], these constants were determined by comparing ODVes1/ODVes2 values in retinal arteries or veins to the mean of retinal SaO2 or SvO2 data, obtained in former studies [18,20]. Here, the ratio ODVes1/ODVes2 plays the same role as the ratio-of-ratios R in pulse oximetry [See Equations [8], [9] and [11]].

One of the two wavelengths in the visible region that are used in retinal oximetry is a wavelength in which the dHb and HbO2 extinction coefficients have a large difference, such as 600, 610 or 555 nm, and the other is 548 or 570 nm [67,68,70], an isosbestic wavelength in the 500–600 nm range [19]. In a number of studies [18,19,20], retinal SO2 was obtained by measuring scattered/reflected light with three wavelengths in the visible region, making use of the method of Pittman and Duling [16], which was mentioned in Section 3.1.

Accuracy and Applications

The accuracy of retinal oximetry can be evaluated by comparing SaO2 values, measured in retinal arteries, to systemic SaO2 values, measured in the peripheral arteries either noninvasively by pulse oximetry or by co-oximeter in extracted arterial blood. The comparison is based on the assumption that SaO2 is constant throughout the entire systemic arterial system, as oxygen diffuses from blood vessels into tissue only in the capillaries.

Comparison of retinal SaO2 readings to systemic SaO2 values was primarily done with two commercial retinal oximeters: The Dynamic Vessel Analyzer [DVA, Imedos, Jena, Germany] and the Oxymap [Oxymap, Reykjavik, Iceland]. The mean of the retinal SaO2 readings by Oxymap, measured in healthy persons, was lower by 5% than finger SaO2, measured by pulse oximetry, with 95% CI of −2 to 12% [71]. Measurements of mean retinal SaO2 in patients with systemic hypoxemia were lower by 2% with 95% CI of −3 to 11% compared to invasive femoral SaO2 values [72]. No significant correlation was found between individual retinal and finger SpO2 [72,73,74]. In a study performed with the DVA retinal oximeter on chronic obstructive pulmonary disease [COPD] patients [75], a low correlation [r = 0.6, p < 0.05] was found between retinal and finger SpO2 readings, and a Bland–Altman plot, comparing retinal and finger SpO2 data, yielded 1.96SD of ±10%. Told et al. [76] found that the individual retinal SaO2 readings, measured by the DVA and Oxymap retinal oximeters, were significantly different.

The lower mean retinal SaO2, relative to finger SpO2, was explained by an error in the process of the retinal oximeters’ calibration or by the countercurrent exchange of oxygen by diffusion from the arteries within the optic nerve [70,76,77]. The low value of retinal SaO2 can also be attributed to the measurement of I0n, the scattered light from the blood vessel neighborhood, which might be affected by light absorption in retinal blood vessels [See Figure 5].

The simultaneous measurement of oxygen saturation in a pair of retinal artery and vein can assess the arteriovenous difference of oxygen saturation and oxygen content, CaO2-CvO2 [See Equation [1]]. The latter provides information on the retinal oxygen consumption in the tissue supplied by the artery and drained by the vein, an important parameter, which is related to the metabolism rate of the retinal cells [78,79]. The arteriovenous difference of oxygen saturation or oxygen content decreased during breathing of 100% oxygen relative to breathing air in measurements on healthy subjects [78,80], but was unchanged in COPD patients [75].

Non-invasive retinal oximetry has been used in physiological studies on healthy persons [68,70,74,78,79,80] and on patients with ophthalmic and a number of systemic diseases [70,75,81,82,83,84,85,86]. The increase in arteriovenous difference of retinal oxygen saturation is generally due to ischemia, while the decrease in that difference might originate from atrophy, which is associated with lower metabolism of retinal cells. Ischemic diseases include diabetes, central retinal vein occlusions, age-related macular degeneration and retinopathy of prematurity. Atrophy-related retinal diseases include glaucoma and retinitis pigmentosa [70].

The accessibility to light of the human retina has been utilized as a window to the brain, as the retina and its optic nerve are considered an extension of the central nervous system, from the anatomic, functional and pathologic aspects [87,88,89]. Like the brain, the retina has a large demand for oxygen and nutrient supplies [88], and both tissues utilize autoregulation to keep sufficient blood supply [90]. Retinal oxygen saturation has been suggested as a mirror to brain oxygenation. In studies on mild cognitive impairment and multiple sclerosis patients [91,92], retinal arteriovenous oxygen saturation difference was significantly lower than that of healthy persons, which reflects reduced oxygen extraction in the retina, indicating reduced metabolism.

5. Near Infrared Spectroscopy [NIRS]

5.1. NIRS Techniques

Near infrared spectroscopy [NIRS] is an oximetric technique that provides assessment of the HbO2 and dHb concentrations [[HbO2] and [dHb], respectively], and of oxygen saturation in the entire blood in tissue—arterial, venous and capillary. The tissue oxygen saturation, StO2, is also named as the tissue oxygenation index, TOI, or regional oxygen saturation, rSO2. Like other oximetric techniques, NIRS is based on the different absorption spectrum of HbO2 and dHb and the measurement of light attenuation in two or more wavelengths. As an optical technique, it can non-invasively measure oxygenation in muscles and cerebral tissue that are accessible to light, since infrared light can penetrate several centimeters in tissue [93,94]. In order to non-invasively assess oxygenation or oxygenation change in the cerebral tissue, light is emitted into the brain through the intact skull and the light transmitted through the cerebral tissue is measured by a detector located 2–4 cm from the light-source. The greater the distance between a pair of optodes [the light-source/detector pair], the deeper cerebral tissue examined. For a typical differential pathlength factor [DPF, the ratio between mean actual pathlength and optodes-distance] value of 5, a distance between optodes of 2–4 cm is equivalent to a transmitted light pathlength of 10–20 cm. Because of the long pathlength, the wavelengths utilized in NIRS are in the near infrared region, 700–1000 nm, where light absorption is relatively small, see Figure 1. Wavelengths above 1000 nm are significantly absorbed by water [95,96].

In the near infrared region, the main absorbers are HbO2 and dHb. The enzyme cytochrome oxidase [CytOx] is also involved in light absorption in cerebral tissue in this wavelength range, but its concentration is much less than the hemoglobin molecules [97]. CytOx is an enzyme in the mitochondrial electron transport chain that synthesizes the adenosine triphosphate [ATP]—the molecule that provides the energy required for cellular metabolism. Since CytOx oxygenated and deoxygenated states differ in their infrared spectrum, NIRS could also be used for the assessment of CytOx oxygenation, but the near-infrared absorption spectra of the two CytOx states overlap those of the hemoglobins, and the concentration of the latter is significantly greater [93,98,99,100,101]. Myoglobin is also involved in infrared light absorption in muscle: its oxygenated and deoxygenated spectra are identical to those of hemoglobin [14,96,102], so that muscle oximetry actually measures the oxygenation state of both hemoglobin and myoglobin molecules [14,103]. The evaluation of CytOx concentration and muscle oxygenation are not discussed in the current review.

As claimed in the current review, the primary problem of oximetry is the discrimination between the absorption and scattering effects in the attenuation measurement. The different light absorption for two wavelengths is the signal, while attenuation due to light-loss G and elongation of the path-length l [Equation [3]] is the primary noise. In arterial pulse oximetry the discrimination between the absorption and scattering effects is achieved by measuring the light attenuation difference between the end-diastole and the end-systole, and assuming that the systolic light attenuation change is mainly due to absorption, neglecting attenuation changes due to the scattering [Equations [7] and [8]]. In NIRS, reflected/transmitted light intensity is measured as a function of time or distance from the light-source, and the diffusion equation, presented below, is utilized for deriving two equations that enable the determination of the two unknowns, the absorption constant µa and the reduced scattering constant µs’. The diffusion equation enables the derivation of the photon fluence rate φ[r,t] that is reflected/transmitted from the medium surface [96,97], provided that the detailed structure and composition of the tissue are given. Since the real tissue structure is complex and unknown, solution of the diffusion equation for a specific tissue is impractical, and a simple model, simulating the real situation, must be used for NIRS measurements.

In general, the model includes light with a narrow cross-section emitted into a semi-infinite homogenous medium that simulates the biological tissue, and a detector that measures the intensity IR[ρ,t] of the reflected light, which is a function of its distance ρ from the light-source and of the time t [if the emitted light is time-dependent]. IR[ρ,t] can be obtained by means of the diffusion equation [96,97,104,105] that relates the rate of change of φ[r,t] to the absorption and scattering coefficients of the medium and the photon emission source intensity.

Solutions satisfying the diffusion equation for the semi-infinite medium model with uniform absorption and scattering coefficients have been determined for three categories of light intensity dependence on time: short impulse, sinusoidally modulated intensity and constant intensity. For each sort of time-dependence, the corresponding solution of the reflected light intensity R[ρ,t] is related to the absorption constant, µa, and the reduced scattering coefficient, µs’. In accordance with the different time-dependences of the illumination, three modalities of NIRS have been developed, sorted as constant intensity [CW] spatially-resolved, time-resolved or frequency-resolved spectroscopy, where each modality makes use of the dependence of R[ρ,t] on the light source-detector distance, temporal dispersion of short light pulse or phase change of high-frequency intensity-modulated light, respectively [96,97,106].

Time-resolved spectroscopy is based on a short impulse emitted into a tissue volume and scattered from the tissue as a broadened pulse, due to different pathlengths between light entrance and exit for different photons. Since the speed of light in tissue is about 2 × 1010 cm/s, the time-of-flight of a photon for typical pathlengths of 2 to 10 cm is 100 to 500 picoseconds. In practice, the short impulse used in time-resolved spectroscopy has a width of a few picoseconds and the broadened reflected pulse demonstrates a time-width of several hundred picoseconds. For the semi-infinite medium model, a solution of the diffusion equation R[ρ,t] can be obtained [96,104,106,107], providing relationship between R[ρ,t] and µ’s and µa. By fitting the theoretical solution to the dependence-on-time of the measured reflected light, µs’ and µa can be determined.

In the frequency-resolved spectroscopy µs’ and µa can also be determined by solving the diffusion equation for the sinusoidal illumination function. The two unknowns, µs’ and µa, can be derived from the amplitude modulation and the phase shift of the reflected light, obtained from the diffusion equation solution [104,106]. Typical values of the modulated light frequency in frequency-resolved spectroscopy are 300–400 MHz.

NIRS measurement of blood oxygenation in the brain is based on light scattering from relatively deep tissue, 1–2 cm below the surface, which necessitates a source-detector distance of 2–4 cm, corresponding to a pathlength of 10–20 cm. Because the light irradiation level into the brain is limited, the transmitted/reflected light has a low intensity and very sensitive detectors must be used. Considering the high time-resolution that is required in the time- and frequency-resolved NIRS for measuring the broadening of the ultrashort pulses or the phase-shift of the very high frequency irradiation, sophisticated techniques of elevated cost must be used. However, sustained technological progress is expected to reduce the costs of the sophisticated time- and frequency-resolved NIRS devices.

In the CW spatially-resolved spectroscopy, reflected light in several wavelengths is measured in several light-sources/detector distances. The solution of the diffusion equation for the semi-infinite medium model relates R[ρ] to some function of the two unknowns, µa and µ’s, and in order to obtain µa separately, several equations have been proposed [96,97,107,108]. By applying the diffusion equation in two wavelengths, two equations for µa[λ] at two wavelengths can be determined and [HbO2] and [dHb] can be derived from the two equations, based on Equation [4]:

εC = µa[λ] = εO[λ][HbO2] + εD[λ][dHb]

[19]

The tissue oxygen saturation, StO2, can be determined from the [HbO2] and [dHb] values, as StO2 = [HbO2]/[[HbO2] + [dHb]].

StO2 can also be obtained by the spatially-resolved NIRS by differentiating the diffusion equation with respect to ρ, which provides a relationship between the gradient of R[ρ] and µaµs’ [96,107,108,109,110]. Since µa = εC

µaµs’ = µs’εC = εOStO2µs’C + εD[1-StO2]µs’C,

[20]

See Equation [19]. Measuring µaµs’ with two wavelengths and assuming that the difference in C between the two wavelengths is negligible and that µs’ decreases slightly and linearly with the wavelength [107,110], StO2 can be derived from Equation [20] in two wavelengths, using an algorithm that is similar to that for SaO2 measurement by pulse oximetry, see Equations [7] and [8].

The accuracy of the NIRS techniques has been challenged on theoretical grounds, as the solution of the diffusion equation is based on a simplistic homogeneous semi-infinite medium model. The model does not reflect the real complex cerebral tissue structure, covered by cerebral-spinal fluid and bone, which also varies between subjects [96,97,111,112]. Fantini et al. [105] also raised an argument against the applicability of the diffusion equation to the semi-infinite model that simulates the tissue in noninvasive NIRS measurements, where the light-source and the detector are placed on the skin surface. The diffusion equation is an approximation that is only appropriate for measurements performed deep inside the bulk medium [105,113].

StO2 was also obtained by means of a single light-source/detector distance, making use of the modified Beer–Lambert equation [Equation [3]] in two or more wavelengths, without utilizing the diffusion equation and the semi-infinite medium model. In that technique, the scattering light-loss G and the absorption by skin blood and pigmentation are not canceled out and the cerebral StO2 is obtained by calibration with saturation measurements in the jugular vein and arterial blood [114,115,116]. The calibration of the cerebral StO2, which is a composite of oxygen saturation in arterial and venous blood, is based on the following assumptions: the venous blood oxygen saturation in a cerebral site is equal to that of the jugular vein, which drains blood from that cerebral site; the saturation in cerebral arterial blood is equal to that of the systemic arterial blood; the arterial blood to venous blood proportion in the cerebral tissue is a known ratio [30:70 or 25:75].

NIRS devices with a single light-source/detector distance are sensitive to the interference of extracerebral blood, while multi-distance NIRS devices can remove contamination from the extra-cranial circulation [117,118]. Some commercial devices only use a single pair of optodes separated by a distance of about 40 mm, which measures absorption in both cerebral and extracranial tissue, but adds measurement at a distance of about 20 mm, which aims to subtract the absorption signal of the extracranial tissue from the former absorption measurement. Measurements of cerebral StO2 by three commercial NIRS devices, with different values of extracerebral tissue perfusion [exploiting scalp ischemia], showed that the spatial resolved technique was not effective in isolating the absorption in the cerebral tissue [119].

Constant intensity illumination with a single light-source/detector distance was also used for measuring changes in tissue oxygenation by making use of the modified Beer–Lambert law [Equation [3]] and avoiding the use of the diffusion equation and the semi-infinite medium model. Temporal changes in [HbO2] and [dHb], induced by functional task, such as respiration or venous or arterial occlusion, can be derived from the corresponding temporal changes in light attenuation [96,97,101,107,120,121], assuming that the effect of light-loss G by scattering is the same for the two states and the attenuation difference between the two states substantially originated from absorption. The process of quantitative assessment of Δ[HbO2] and Δ[dHb] from the difference between the transmitted/reflected light intensity in the two states, I1 and I2, is similar to that of SaO2 measurement by pulse oximetry [See Equations [7] and [8]]:

ln[I1/I2] = εΔCl = [εOΔ[HbO2] + εDΔ[dHb]]l.

[21]

Measurement of ln[I1/I2] in two wavelengths provides two equations with two unknowns, Δ[HbO2] and Δ[dHb], that can be solved for the two unknowns. The pathlength l for the two wavelengths can be obtained from the distance between the two optodes and the differential pathlength factor DPF. The DPF has been determined from time-resolved NIRS or by measuring the mean time-of-flight of ultrashort pulses of light propagating through the tissue [96,97,107,112,122], but since the variability among patients is large [96,120], the measurement accuracy is low. Nevertheless, qualitative information on temporal changes of [HbO2] and [dHb] due to functional tasks is commonly used in functional NIRS [fNIRS].

Cerebral fNIRS measures temporal changes of infrared light intensity transmitted through the brain, yielding temporal changes of [HbO2] and [dHb] in the cerebral tissue that follow functional tasks. The latter is originated from cerebral neural activity that results in a local increase in tissue metabolism and oxygen consumption, followed by vasodilatation and increased blood flow [neurovascular coupling] [123,124,125,126,127].

In cerebral fNIRS, one or a number of non-invasive skull probes are used to evaluate the increase in brain activity at specific locations in the cerebral cortex. Functional near infrared imaging [fNIRI] refers to the simultaneous measurement of local temporal oxygenation changes in multiple sites in a large area. In fNIRS and fNIRI, CW NIRS devices are commonly used, due to their small size, suitable time resolution and low-cost, despite the inability of the technique to quantify changes in [HbO2] and [dHb] in absolute terms. The determination of the brain activity location by fNIRS and fNIRI is of low spatial resolution [as compared to other brain imaging modalities, such as fMRI], because of the strong scattering of light in tissue and the long pathlength in NIRS [125,126].

Multi-channel NIRS, followed by image reconstruction, provides a two-dimensional image of the cerebral oxygenation level. A further advancement of fNIRI is the three-dimensional presentation of the cerebral oxygenation—diffuse optical tomography [DOT], which includes the ability to measure light after transmission through several overlapping light-source/detector distances, in each of the multiple sensors. The assessment of the oxygenation dependence on tissue depth is based on the fact that greater distance between the optodes is associated with deeper tissue pathlengths [125,126,127]. The limited number of light-sources and detectors sets a trade-off between resolution and measurement region area. DOT measurements on the whole cerebral cortex, with relatively high resolution, have been performed on newborns and infants, primarily for neurodevelopmental assessment and detection of neurodevelopmental disorders [126,128,129,130].

5.2. Validation and Accuracy

At present, the primary implementation of NIRS, in both clinical diagnosis and research, is in the assessment of tissue oxygenation in cerebral tissue, which is directly related to cerebral blood perfusion [131,132,133,134]. Timely assessment of the latter is important in critically ill patients and patients undergoing major surgical operations, because of their increased incidence of adverse outcomes, particularly neurological complications that may occur due to cerebral hypo-perfusion. Hence, a reliable noninvasive technique for StO2 assessment is expected to be beneficial for the improvement of patient treatment and outcomes in perioperative and intensive care medicine.

A major difficulty of validating the cerebral oxygenation NIRS technology is the absence of acceptable and reliable methods for the measurement of cerebral tissue oxygenation to compare NIRS with [135]. In several studies, NIRS-derived cerebral StO2 readings were compared to oxygen saturation in the jugular vein [SjO2], which drains venous blood from one of the brain hemispheres. Both NIRS StO2 and invasive SjO2 decreased during induced hypoxia, though the deviations between the values of the two parameters were significant [100,135]. In the study of Rosenthal et al. [136], NIRS StO2 correlated with SjO2 when the latter was measured in blood drained from the hemisphere with the NIRS probe site [r = 0.60, p < 0.001], but not with SjO2 measured in the jugular vein that drains blood from the contralateral hemisphere.

As StO2 is the mean oxygen saturation in the entire blood in tissue, arterial and venous, cerebral StO2 readings were compared to a weighted mean of oxygen saturation in blood extracted from the jugular vein and a systemic artery [SavO2] [114,115,135], and the comparison rendered smaller deviations than those found for SjO2. The weight of the cerebral arterial and venous blood is generally based on the assumption that the venous-to-arterial blood volume ratio of the cerebral tissue is 70:30, as was found by PET measurements [137]:

SavO2 = 0.3SaO2 + 0.7SjO2.

[22]

SaO2 is generally measured by pulse oximetry or by extracting blood from an artery utilizing a co-oximeter; SjO2 is measured invasively by means of a co-oximeter in extracted blood from a jugular vein.

Validation studies of commercial cerebral oximeters were performed on healthy volunteers or patients with jugular catheterization, after calibrating NIRS StO2, by invasive measurements of SavO2 on a different group of healthy volunteers. In these studies [114,115,116,138,139,140], the StO2/SavO2 correlation coefficient values were 0.7–0.9, the bias values [mean difference between their individual values] were 0.5–1.2%, and the standard deviation values were 3–5.4%.

Bickler et al. [100] compared cerebral StO2, measured in healthy volunteers by five commercial cerebral oximeters, with invasive SavO2 during induced hypoxemia. The mean bias values for the different devices were −1.13–2.84 and the standard deviation was 3.92 for one of the devices and 6.27–9.72 for the others. While the cerebral oximeters were found to be effective in detecting the induced desaturations, the significant variation in baseline readings among the five devices demonstrated inaccuracy in absolute cerebral StO2 measurement by the current NIRS devices. Furthermore, the great variation in cerebral StO2 values among the healthy subjects limits the ability of the current non-invasive cerebral oximeters to yield a definite threshold value of the cerebral StO2 that might lead to tissue damage [100].

La Cour et al. [135] reviewed available in vivo validation studies, which compared noninvasive cerebral NIRS-based StO2 to SavO2 in pediatric and adult populations, and found poor agreement in both adults and children: the difference between the limits-of-agreement values was of about 13%. [The limits-of-agreement are the mean of the differences between the two parameters ±1.96 standard-deviations]. A comparison of StO2 to SjO2 rendered lower agreement: the difference between the limits-of-agreement was about 18%. The poor agreement between cerebral StO2 and SavO2 can be attributed to differences of the arterial-to-venous blood-volume ratio between subjects and differences within subjects, due to changes in hypoxia level. In addition, the jugular vein drains blood from the whole brain hemisphere, while the non-invasive NIRS only measures saturation in the cerebral tissue, and StO2 in the cerebral and non-cerebral tissues are not necessarily equal [100,135].

The partial pressure of oxygen in a tissue [PtO2] is a physiological parameter that is related to StO2 through the oxygen–hemoglobin dissociation curve. Cerebral PtO2 is measured invasively in the brain tissue of patients with traumatic brain injury for clinical purposes, and in a number of studies, its association with cerebral StO2 was examined. Despite the relationship of both parameters to the balance between tissue oxygen delivery and consumption, only poor or no correlation between cerebral StO2 and PtO2 was found [117,118,141,142], probably because of the nonlinear relationship between StO2 and PtO2 in the blood [the oxygen–hemoglobin dissociation curve].

The precision of NIRS devices, i.e., the variation of repeated measurements at the same site, was also evaluated in a number of studies on neonates, rendering values of 5.2% [111] and 2.0% on term and 4.2% on preterm neonates [143]. In a recent study [144] on preterm neonates, a precision value of 2.64% was found, but some part of the StO2 variability was attributed to spontaneous hemodynamic fluctuations. Using methods for reducing the fluctuations’ effect, an improved precision value of 1.85% was achieved.

The validation studies dealt with in the current subsection, and evaluated the accuracy of cerebral StO2 readings measured in absolute terms by NIRS cerebral oximeters, based on CW spatially-resolved spectroscopy and calibration by invasive SjO2 measurements. In those studies, the accuracy of the commercial devices based on the NIRS technique was found to be low, and the inter-subject and inter-device deviations were substantial. Nevertheless, qualitative information on cerebral StO2 can be obtained by NIRS, and changes in cerebral StO2 as a function of time can be detected reliably, as will be described in the following subsection.

5.3. Clinical Applications

The potential benefit of NIRS cerebral StO2 readings for clinical treatment has been evaluated in a number of studies, in particular in randomized intervention studies on patients undergoing major surgical operations. In the latter studies, in the intervention group, cerebral StO2 was monitored during the surgical operation by NIRS, and supplemental oxygen dosage was adjusted according to StO2 level, while in the control group, the treatment was not directed by NIRS readings. In several studies, the intervention group demonstrated a decreased incidence of adverse events compared with the control group [145,146,147], while in similar studies [148,149,150], the NIRS-guided intervention was not associated with improved patient outcomes. Yu et al. [151], in a systematic review, inferred low quality of evidence for the beneficial effect of cerebral NIRS monitoring on postoperative adverse events, such as stroke, delirium or death, due to a low number of events and wide confidence intervals.

The benefit of NIRS cerebral StO2 measurements for clinical treatment in preterm neonates has also been evaluated in a randomized intervention clinical trial, performed in eight neonatal intensive care units [152,153,154]. In the intervention group, the burden of hypoxia was significantly lower than that in the control group, but no decrease was found in all-cause mortality, brain injury score and other clinical outcomes [152,153]. No difference in neurodevelopment outcomes at two years of corrected age was found between the two groups [154].

In observational studies, decreased perioperative cerebral StO2 was associated with poor patient outcomes after a major surgery [155,156]. In a recent systematic review [133], an association between low cerebral StO2 and higher incidence of delirium was found in four studies in critically ill patients, compared to controls [157,158,159,160], but only in two of them [159,160] was the relationship statistically significant.

Continuous assessment of brain tissue blood flow and oxygenation of brain tissue is clinically important for traumatic brain injury [TBI] patients, for whom rapid diagnosis is essential for the timely restoration of brain metabolic function and the avoidance of secondary brain injury. In order to identify episodes of cerebral ischemia and hypoxia that can deteriorate patients’ clinical state, several invasive techniques are being used to measure physiological markers, such as intracranial pressure and PtO2. The opportunity to replace those invasive modalities with noninvasive ones has motivated observational studies that examined the association of CW NIRS StO2 monitoring in adult TBI patients with their functional outcome and with physiological-neurological parameters [117,118,161]. Several studies showed that the detection and evaluation of cerebral hypoxia by NIRS were related to increased mortality or poor functional outcome, but the evidence was relatively weak, and additional research is needed before NIRS-based techniques are routinely utilized as a reliable clinical tool.

An interesting application of NIRS StO2 monitoring is the assessment of cerebral autoregulation [CAR]—the ability of the brain to keep cerebral blood flow [CBF] relatively constant during changes of mean arterial blood pressure, as long as the latter is maintained within defined limits [Figure 5] [162,163,164]. CAR protects the brain from hypo-perfusion caused by low perfusion-pressure and from hypertension episodes that can lead to elevated intracranial pressure and to damage to the micro-vessels. In adverse events such as traumatic brain injury, the CAR mechanism might be compromised, potentially leading to an exacerbation of the patient’s condition. Evaluation of CAR can be done from the relationship between the temporal changes in CBF and those in mean arterial blood pressure [MAP]—a positive correlation between them indicates impaired CAR, while intact autoregulation leads to a low correlation [118]. However, since CBF cannot be measured directly, surrogate parameters are being used, such as intra-cranial pressure, measured invasively, or trans-cranial Doppler velocity, measured in the middle cerebral artery. A number of quantitative CAR indices, based on those surrogate measurements of CBF, have been developed for CAR assessment [165]. Cerebral NIRS-derived StO2 has also been used as a surrogate for spontaneous CBF changes, as tissue oxygenation is directly related to tissue perfusion, and continuous NIRS StO2 measurement is simpler to use than continuous trans-cranial Doppler measurement.

NIRS-derived CAR indices have been compared to those based on trans-cranial Doppler monitoring, and correlation values of 0.40–0.81 were found between indices obtained by the two methods [117,118,166,167,168,169,170]. In a study performed on children during cardiopulmonary bypass [171], high values of the NIRS-derived CAR index were associated with hypotension. The latter study also found that the lower MAP limit of the CAR curve [Figure 5] can be determined from the NIRS CAR index.

The correlation between the two sets of CAR indices, obtained by trans-cranial Doppler velocimetry and by NIRS, is significant but low, and only allows validation in qualitative terms of the NIRS-based CAR assessment. As was described above, validation studies of the cerebral oximeters reported low accuracy of the technique and substantial inter-subject and inter-device deviations when StO2 readings were measured in absolute terms, while changes of cerebral StO2 can be obtained qualitatively.

The review was focused on oximetric measurements on humans, but cerebral oxygenation has also been measured by NIRS and fNIRS on animals, both for physiological and pathological animal research and for the assessment of the techniques. Some references [172,173,174,175,176], not representative of studies describing cerebral NIRS and fNIRS measurements on animals, were presented.

6. Conclusions

Due to the paramount physiological importance of adequate oxygen supply to the tissue, clinical evaluation of blood oxygenation in arteries and veins is essential, motivating the development of clinical techniques for the measurement and monitoring of blood oxygenation in arteries, in veins or in both arteries and veins in the tissue. The non-invasive nature of light propagation through tissue and the different spectra of oxygenated and deoxygenated hemoglobin stimulated the development of several oximetric techniques for the assessment of oxygen saturation and concentrations of oxygenated and deoxygenated hemoglobin, parameters that provide information on blood oxygenation. The current review presents the main oximetric techniques for the assessment of blood oxygenation by measurements of attenuation in tissue of light comprising two wavelengths, utilizing the different absorption spectra of oxygenated and deoxygenated hemoglobin. The various oximetric techniques enable the evaluation of oxygenation in arterial or venous blood or in the entire blood in tissue, in finger, peripheral muscle, brain or retina—tissues that are accessible to light in the visible or infrared regions.

The primary challenge of oximetry is the discrimination between the effects of absorption by hemoglobin [signal] and scattering by tissue elements [noise] in the attenuation process, and several mechanisms have been created to isolate the absorption effect. Currently, only two techniques, pulse oximetry for the measurement of oxygen saturation in arterial blood, and NIRS, for the assessment of tissue oxygenation in cerebral and muscle tissue, are being used in clinical practice, and only pulse oximetry has been widely accepted as a valid diagnostic tool. The use of photoplethysmography, for the isolation of absorption in arterial blood, enabled pulse oximetry to provide arterial blood oxygen saturation measurements with reasonable accuracy. The accuracy of pulse oximetry seems to be sufficient for most clinical requirements, though in patients under oxygen supplementation, preterm neonates in particular, greater accuracy is required to avoid excessive oxygenation. In regard to NIRS, it seems that significant improvement in its technical aspects is required in order to obtain reliable absolute measurement of cerebral oxygen saturation, a prerequisite for a medical device to be widely accepted as a clinical tool.

Author Contributions

M.N. conceptualized and wrote the review. I.N. and Y.A. critically reviewed it. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflict of interest.

Abbreviations

CaO2 oxygen content in arterial blood
[Hb] hemoglobin concentration in blood
HbO2 oxygenated hemoglobin
dHb deoxygenated hemoglobin
SO2 oxygen saturation in blood vessels, arteries or veins
SaO2 arterial oxygen saturation
SpO2 arterial oxygen saturation measured by pulse oximetry
PaO2 arterial oxygen tension
SvO2 oxygen saturation in venous blood
StO2 oxygen saturation in tissue blood
SjO2 oxygen saturation in the jugular vein
SavO2 weighted mean of oxygen saturation in blood in the jugular vein and in a systemic artery
µa absorption constant
µs scattering coefficient
µs’ reduced scattering coefficient
g anisotropy
εO extinction coefficients of HbO2
εD extinction coefficients of dHb
ε the mean extinction coefficient of blood in the tissue vessels, arteries and veins
C hemoglobin concentration
I0 incident light intensity
It transmitted light intensity
d emitter-detector distance
l mean optical path-length
DPF differential pathlength factor
G light-loss due to tissue scattering
OD optical density
PPG photoplethysmography
ID The maximal value of the PPG pulse
IS The minimal value of the PPG pulse
AC the difference between the maximal and minimal values of the PPG pulse
DC the mean light transmission during the pulse
λ wavelengths
ΔC the maximal hemoglobin concentration increase in the tissue during systole
R ratio of ratios
NIRS Near infrared spectroscopy
TOI tissue oxygenation index
rSO2 regional oxygen saturation
CytOx cytochrome oxidase
IR[ρ,t] the intensity of the transmitted/reflected light in NIRS
ρ the distance from the light-source
t time
CW constant wave/intensity
fNIRS functional NIRS
fNIRI functional near infrared imaging
DOT diffuse optical tomography
PtO2 oxygen partial pressure in a tissue
TBI traumatic brain injury
CAR cerebral autoregulation
CBF cerebral blood flow
MAP mean arterial blood pressure

References

  1. Plötz, F.B.; van Lingen, R.A.; Bos, A.P. Venous oxygen measurements in the inferior vena cava in neonates with respiratory failure. Crit. Care 1998, 2, 57–60. [Google Scholar] [CrossRef] [PubMed][Green Version]
  2. Osypka, M.; Soleymani, S.; Seri, I.; Noori, S. Assessment of Cardiac output in neonates: Techniques using the Fick principle, pulse wave analysis and electrical impedance. In Hemodynamics and Cardiology: Neonatology Questions and Controversies, 2nd ed.; Kleinman, C.S., Seri, I., Eds.; Elsevier Saunders: Philadelphia, PA, USA, 2012; Chapter 6; pp. 125–149. [Google Scholar]
  3. Prahl, S. The absorption spectra of the oxygenated and deoxygenated hemoglobin molecules. Available online: //omlc.ogi.edu/spectra/hemoglobin/index.html [accessed on 21 January 2014].
  4. Zuzak, K.J.; Schaeberle, M.D.; Gladwin, M.T.; Cannon, R.O., 3rd; Levin, I.W. Noninvasive Determination of Spatially Resolved and Time-Resolved Tissue Perfusion in Humans During Nitric Oxide Inhibition and Inhalation by Use of a Visible-Reflectance Hyperspectral Imaging Technique. Circulation 2001, 104, 2905–2910. [Google Scholar] [CrossRef] [PubMed][Green Version]
  5. Greenman, R.L.; Panasyuk, S.; Wang, X.; Lyons, T.E.; Dinh, T.; Longoria, L.; Giurini, J.M.; Freeman, J.; Khaodhiar, L.; Veves, A. Early Changes in the Skin Microcirculation and Muscle Metabolism of the Diabetic Foot. Lancet 2005, 366, 1711–1717. [Google Scholar] [CrossRef]
  6. Benaron, D.A.; Parachikov, I.H.; Cheong, W.F.; Friedland, S.; Rubinsky, B.E.; Otten, D.M.; Liu, F.W.H.; Levinson, C.J.; Murphy, A.L.; Price, J.W.; et al. Design of a visible-light spectroscopy clinical tissue oximeter. J. Biomed. Opt. 2005, 10, 44005. [Google Scholar] [CrossRef]
  7. Van Noord, D.; Sana, A.; Benaron, D.A.; Pattynama, P.M.T.; Verhagen, H.J.M.; Hansen, B.E.; Kuipers, E.J.; Mensink, P.B.F. Endoscopic visible light spectroscopy: A new, minimally invasive technique to diagnose chronic GI ischemia. Gastrointest. Endosc. 2011, 73, 291–298. [Google Scholar] [CrossRef]
  8. Jacques, S.L. Optical properties of biological tissues: A review. Phys. Med. Biol. 2013, 58, R37–R61. [Google Scholar] [CrossRef]
  9. Jonasson, H.; Fredriksson, I.; Bergstrand, S.; Östgren, C.J.; Larsson, M.; Strömberg, T. In vivo characterization of light scattering properties of human skin in the 475- to 850-nm wavelength range in a Swedish cohort. J. Biomed. Opt. 2018, 23, 121608. [Google Scholar] [CrossRef][Green Version]
  10. Friebel, M.; Helfmann, J.; Netz, U.; Meinke, M. Influence of oxygen saturation on the optical scattering properties of human red blood cells in the spectral range 250 to 2000 nm. J. Biomed. Opt. 2009, 14, 034001. [Google Scholar] [CrossRef]
  11. Jacques, S.L. Time-resolved reflectance spectroscopy in turbid tissues. IEEE Trans. Biomed. Eng. 1989, 36, 1155–1161. [Google Scholar] [CrossRef]
  12. Matcher, S.J.; Cooper, C.E. Absolute quantification of deoxyhaemoglobin concentration in tissue near infrared spectroscopy. Phys. Med. Biol. 1994, 39, 1295–1312. [Google Scholar] [CrossRef]
  13. Zijlstra, W.G.; Buursma, A.; Meeuwsen-van der Roest, W.P. Absorption Spectra of Human Fetal and Adult Oxyhemoglobin, De-Oxyhemoglobin, Carboxyhemoglobin, and Methemoglobin. Clin. Chem. 1991, 37, 1633–1638. [Google Scholar] [CrossRef] [PubMed]
  14. Barstow, T.J. Understanding near infrared spectroscopy and its application to skeletal muscle research. J. Appl. Physiol. 2019, 126, 1360–1376. [Google Scholar] [CrossRef] [PubMed]
  15. Twersky, V. Interface effects in multiple scattering by large, low-refracting, absorbing particles. J. Opt. Soc. Am. 1970, 60, 908–914. [Google Scholar] [CrossRef]
  16. Pittman, R.N.; Duling, B.R. A new method for the measurement of percent oxyhemoglobin. J. Appl. Physiol. 1975, 38, 315–320. [Google Scholar] [CrossRef]
  17. Iwasaki, H.; Yoshizaki, K.; Koyano, H. Fiber-optic transmission catheter for regional venous oxygen saturation and blood flow. J. Appl. Physiol. 1992, 72, 1616–1621. [Google Scholar] [CrossRef]
  18. Delori, F.C. Noninvasive technique for oximetry of blood in retinal vessels. Appl. Opt. 1988, 27, 1113–1125. [Google Scholar] [CrossRef]
  19. Smith, M.H. Optimum wavelength combinations for retinal vessel oximetry. Appl. Opt. 1999, 38, 258–267. [Google Scholar] [CrossRef]
  20. Harris, A.; Dinn, R.B.; Kagemann, L.; Rechtman, E. A review of methods for human retinal oximetry. Ophthalmic. Surg. Lasers Imaging 2003, 34, 152–164. [Google Scholar] [CrossRef]
  21. Millikan, G.A. The oximeter: An instrument for measuring continuously oxygen saturation of arterial blood in man. Rev. Sci. Instrum. 1942, 13, 434–444. [Google Scholar] [CrossRef]
  22. Merrick, E.B.; Hayes, T.J. Continuous, Non-Invasive Measurements of Arterial Blood Oxygen Levels. Hewlett-Packard J. 1976, 28, 2–9. Available online: //archive.org/details/Hewlett-Packard_Journal_Vol._28_No._2_1976-10_Hewlett-Packard/page/n1/mode/2up [accessed on 26 August 2020].
  23. Saunders, N.A.; Powels, A.C.P.; Rebuck, A.S. Ear Oximetry: Accuracy and Practicability in the Assessment of Arterial Oxygenation. Am. Rev. Respir. Dis. 1976, 113, 745–749. [Google Scholar]
  24. Douglas, N.J.; Brash, H.M.; Wraight, P.K.; Calverley, P.M.A.; Legget, R.J.E.; McElderry, L.; Flenley, D.C. Accuracy, sensitivity to carboxyhemoglobin and speed of response of the Hewlett Packard 47201A ear oximeter. Am. Rev. Respir. Dis. 1979, 119, 311–313. [Google Scholar] [PubMed]
  25. West, P.; George, C.F.; Kryger, M.H. Dynamic in vivo response characteristics of three oximeters: Hewlett-Packard 47201A, Biox III, and Nellcor N-100. Sleep 1987, 10, 263–271. [Google Scholar] [CrossRef] [PubMed]
  26. Severinghaus, J.W. Takuo Aoyagi: Discovery of pulse oximetry. Anesth. Analg. 2007, 105, S1–S4. [Google Scholar] [CrossRef] [PubMed]
  27. Molnar, Z.; Umgelter, A.; Toth, I.; Livingstone, D.; Weyland, A.; Sakka, S.G.; Meier-Hellmann, A. Continuous Monitoring of ScvO[2] by a New Fibre-Optic Technology Compared With Blood Gas Oximetry in Critically Ill Patients: A Multicentre Study. Intensive Care Med. 2007, 33, 1767–1770. [Google Scholar] [CrossRef]
  28. Iodice, F.G.; Ricci, Z.; Haiberger, R.; Favia, I.; Cogo, P. Fiberoptic monitoring of central venous oxygen saturation [PediaSat] in small children undergoing cardiac surgery: Continuous is not continuous. F1000Research 2014, 3, 23. [Google Scholar] [CrossRef]
  29. Herner, A.; Haller, B.; Mayr, U.; Rasch, S.; Offman, L.; Schmid, R.; Huber, W. Accuracy and precision of SCVO2 measured with the CeVOX device: A prospective study in patients with a wide variation of SCVO2-values. PLoS ONE 2018, 13, e0192073. [Google Scholar] [CrossRef]
  30. Baulig, W.; Spielmann, N.; Zaiter, H.; Lijovic, T.; Bettex, D.; Bürki, C.; Weiss, M. In-vitro evaluation of the PediaSat continuous central venous oxygenation monitoring system. Eur. J. Anaesthesiol. 2010, 27, 289–294. [Google Scholar] [CrossRef][Green Version]
  31. Stocchetti, N.; Magnoni, S.; Zanier, E.R. My paper 20 years later: Cerebral venous oxygen saturation studied with bilateral samples in the internal jugular veins. Intensive Care Med. 2015, 41, 412–417. [Google Scholar] [CrossRef]
  32. Gupta, A.K.; Hutchinson, P.J.; Al-Rawi, P.; Gupta, S.; Swart, M.; Kirkpatrick, P.J.; Menon, D.K.; Datta, A.K. Measuring brain tissue oxygenation compared with jugular venous oxygen saturation for monitoring cerebral oxygenation after traumatic brain injury. Anesth. Analg. 1999, 88, 549–553. [Google Scholar] [CrossRef]
  33. Gopinath, S.P.; Robertson, C.S.; Contant, C.F.; Hayes, C.; Feldman, Z.; Narayan, R.K.; Grossman, R.G. Jugular venous desaturation and outcome after head injury. Neurol. Neurosurg. Psychiatr. 1994, 57, 717–723. [Google Scholar] [CrossRef][Green Version]
  34. White, H.; Baker, A. Continuous jugular venous oximetry in the neurointensive care unit—A brief review. Can. J. Anaesth. 2002, 49, 623–629. [Google Scholar] [CrossRef] [PubMed]
  35. Schoon, P.; Benito Mori, L.; Orlandi, G.; Larralde, C.; Radrizzan, M. Incidence of intracranial hypertension related to jugular bulb oxygen saturation disturbances in severe traumatic brain injury patients. Acta Neurochir. 2002, 81, 285–287. [Google Scholar]
  36. Yoshiya, I.; Shimada, Y.; Tanake, K. Spectrophotometric monitoring of arterial oxygen saturation on the fingertip. Med. Biol. Eng. Comput. 1980, 18, 27–32. [Google Scholar] [CrossRef] [PubMed]
  37. Wieben, O. Light absorbance in pulse oximetry. In Design of Pulse Oximeters; Webster, J.G., Ed.; Institute of Physics Publishing: Bristol, UK, 1997; pp. 40–55. ISBN 9780750304672. [Google Scholar]
  38. Mannheimer, P.D.; Casciani, J.R.; Fein, M.E.; Nierlich, S.L. Wavelength selection for low-saturation pulse oximetry. IEEE Trans. Biomed. Eng. 1997, 44, 148–158. [Google Scholar] [CrossRef] [PubMed]
  39. Nitzan, M.; Babchenko, A.; Khanokh, B.; Taitelbaum, H. The measurement of oxygen saturation in venous blood by dynamic near IR spectroscopy. J. Biomed. Opt. 2000, 5, 155–162. [Google Scholar] [CrossRef] [PubMed][Green Version]
  40. Kyriacou, P.A. Pulse oximetry in the esophagus. Physiol. Meas. 2006, 27, R1–R35. [Google Scholar] [CrossRef]
  41. Yossef Hay, O.; Cohen, M.; Nitzan, I.; Kasirer, Y.; Shahroor-Karni, S.; Yitzhaky, Y.; Engelberg, S.; Nitzan, M. Pulse Oximetry with Two Infrared Wavelengths without Calibration in Extracted Arterial Blood. Sensors 2018, 18, 3457. [Google Scholar] [CrossRef][Green Version]
  42. Kim, J.G.; Liu, H. Variation of haemoglobin extinction coefficients can cause errors in the determination of haemoglobin concentration measured by near-infrared spectroscopy. Phys. Med. Biol. 2007, 52, 6295–6322. [Google Scholar] [CrossRef]
  43. Kim, J.G.; Liu, H.; [Joint Graduate Program in Biomedical Engineering, University of Texas at Arlington/University of Texas Southwestern Medical Center at Dallas, Arlington, TX 7601, USA]. Personal communication, 2007.
  44. Zourabian, A.; Siegel, A.; Chance, B.; Ramanujan, N.; Rode, M.; Boas, D.A. Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse oximetry. J. Biomed. Opt. 2000, 5, 391–405. [Google Scholar] [CrossRef][Green Version]
  45. Schowalter, J.S. Calibration. In Design of Pulse Oximeters; Webster, J.G., Ed.; Institute of Physics Publishing: Bristol, UK, 1997; Chapter 10. [Google Scholar]
  46. Hornberger, C.; Wabnitz, H. Approaches for calibration and validation of near-infrared optical methods for oxygenation monitoring. Biomed. Tech. 2018, 25, 537–546. [Google Scholar] [CrossRef]
  47. Seguin, P.; Le Rouzo, A.; Tanguy, M.; Guillou, Y.M.; Feuillu, A.; Mallédant, Y. Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. Crit. Care Med. 2000, 28, 703–706. [Google Scholar] [CrossRef]
  48. Van de Louw, A.; Cracco, C.; Cerf, C.; Harf, A.; Duvaldestin, P.; Lemaire, F.; Brochard, L. Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med. 2001, 27, 1606–1613. [Google Scholar] [CrossRef] [PubMed]
  49. Wilson, B.J.; Cowan, H.J.; Lord, J.A.; Zuege, D.J.; Zygun, D.A. The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: A retrospective cohort study. BMC Emerg. Med. 2010, 10, 9. [Google Scholar] [CrossRef] [PubMed][Green Version]
  50. Rosychuk, R.J.; Hudson-Mason, A.; Eklund, D.; Lacaze-Masmonteil, T. Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in very preterm infants. Neonatology 2012, 101, 14–19. [Google Scholar] [CrossRef] [PubMed]
  51. Ross, P.A.; Newth, C.J.; Khemani, R.G. Accuracy of pulse oximetry in children. Pediatrics 2014, 133, 22–29. [Google Scholar] [CrossRef][Green Version]
  52. Lakshminrusimha, S.; Manja, V.; Mathew, B.; Suresh, G.K. Oxygen targeting in preterm infants: A physiological interpretation. J. Perinatol. 2015, 35, 8–15. [Google Scholar] [CrossRef][Green Version]
  53. Cummings, J.J.; Polin, R.A. Committee on fetus and newborn. Oxygen targeting in extremely low birth weight infants. Pediatrics 2016, 138, e20161576. [Google Scholar] [CrossRef][Green Version]
  54. Thijssen, M.; Janssen, L.; le Noble, J.; Foudraine, N. Facing SpO2 and SaO2 discrepancies in ICU patients: Is the perfusion index helpful? J. Clin. Monit. Comput. 2019. [Google Scholar] [CrossRef][Green Version]
  55. SUPPORT Study Group. Target Ranges of Oxygen Saturation in Extremely Preterm Infants. N. Engl. J. Med. 2010, 362, 1959–1969. [Google Scholar] [CrossRef][Green Version]
  56. BOOST-II Australia and United Kingdom Collaborative Groups; Tarnow-Mordi, W.; Stenson, B.; Kirby, A.; Juszczak, E.; Donoghoe, M.; Deshpande, S.; Morley, C.; King, A.; Doyle, L.W.; et al. Outcomes of Two Trials of Oxygen-Saturation Targets in Preterm Infants. N. Engl. J. Med. 2016, 374, 749–760. [Google Scholar]
  57. Hafner, S.; Beloncle, F.; Koch, A.; Radermacher, P.; Asfar, A. Hyperoxia in intensive care, emergency, and peri-operative medicine: Dr. Jekyll or Mr. Hyde? A 2015 update. Ann. Intensive Care 2015, 5, 42. [Google Scholar] [CrossRef] [PubMed][Green Version]
  58. Demiselle, J.; Wepler, M.; Hartmann, C.; Radermacher, P.; Schortgen, F.; Meziani, F.; Singer, M.; Seegers, V.; Asfar, P. Hyperoxia toxicity in septic shock patients according to the Sepsis-3 criteria: A post hoc analysis of the HYPER2S trial. Ann. Intensive Care 2018, 8, 90. [Google Scholar] [CrossRef] [PubMed][Green Version]
  59. Nitzan, M.; Noach, S.; Tobal, E.; Adar, Y.; Miller, Y.; Shalom, E.; Engelberg, S. Calibration-free pulse oximetry based on two wavelengths in the infrared—A preliminary study. Sensors 2014, 14, 7420–7434. [Google Scholar] [CrossRef] [PubMed]
  60. Khan, M.; Pretty, C.G.; Amies, A.C.; Balmer, J.; Banna, H.E.; Shaw, G.M.; Geoffrey Chase, J. Proof of concept non-invasive estimation of peripheral venous oxygen saturation. Biomed. Eng. Online 2017, 16, 60. [Google Scholar] [CrossRef] [PubMed][Green Version]
  61. Wolf, M.; Due, G.; Keel, M.; Niederer, P.; von Siebenthal, K.; Bucher, H.-U. Continuous noninvasive measurement of cerebral arterial and venous oxygen saturation at the bedside in mechanically ventilated neonates. Crit. Care Med. 1997, 25, 1579–1582. [Google Scholar] [CrossRef]
  62. Walton, Z.D.; Kyriacou, P.A.; Silverman, D.G.; Shelley, K.H. Measuring venous oxygenation using the photoplethysmograph waveform. J. Clin. Monit. Comput. 2010, 24, 295–303. [Google Scholar] [CrossRef]
  63. Duadi, H.; Nitzan, M.; Fixler, D. Simulation of oxygen saturation measurement in a single blood vein. Opt. Lett. 2016, 41, 4312–4315. [Google Scholar] [CrossRef]
  64. Hammer, M.; Leistritz, S.; Leistritz, L.; Schweitzer, D. Light Paths in Retinal Vessel Oxymetry. IEEE Trans. Biomed. Eng. 2001, 48, 592–598. [Google Scholar] [CrossRef]
  65. Beach, J.M.; Schwenzer, K.J.; Srinivas, S.; Kim, D.; Tiedeman, J.S. Oximetry of retinal vessels by dual wavelength imaging: Calibration and influence of pigmentation. J. Appl. Physiol. 1999, 86, 748–758. [Google Scholar] [CrossRef][Green Version]
  66. Hardarson, S.H.; Harris, A.; Karlsson, R.; Halldorsson, G.; Kagemann, L.; Rechtman, E.; Zoega, G.M.; Eysteinsson, T.; Benediktsson, J.A.; Thorsteinsson, A.; et al. Automatic Retinal Oximetry. Investig. Ophthalmol. Visual Sci. 2006, 47, 11. [Google Scholar] [CrossRef]
  67. Hammer, M.; Vilser, W.; Riemer, T.; Schweitzer, D. Retinal Vessel Oximetry-Calibration, Compensation for Vessel Diameter and Fundus Pigmentation, and Reproducibility. J. Biomed. Opt. 2008, 13, 054015. [Google Scholar] [CrossRef] [PubMed][Green Version]
  68. Hammer, M.; Vilser, W.; Riemer, T.; Liemt, F.; Jentsch, S.; Dawczynski, J.; Schweitzer, D. Retinal Venous Oxygen Saturation Increases by Flicker Light Stimulation. IOVS 2011, 52, 1. [Google Scholar] [CrossRef][Green Version]
  69. Mordant, D.J.; Al-Abboud, I.; Muyo, G.; Gorman, A.; Sallam, A.; Ritchie, P.; Harvey, A.R.; McNaught, A.I. Spectral imaging of the retina. Eye 2011, 25, 309–320. [Google Scholar] [CrossRef] [PubMed][Green Version]
  70. Stefánsson, E.; Olafsdottir, B.O.; Eliasdottir, T.S.; Wehmeijer, W.; Einarsdottir, A.B.; Bek, T.; Torp, T.L.; Grauslund, J.; Eysteinsson, T.; Karlsson, R.A.; et al. Retinal oximetry: Metabolic imaging for diseases of the retina and brain. Progress Retinal Eye Res. 2019, 70, 1–22. [Google Scholar] [CrossRef] [PubMed]
  71. Traustason, S.; Jensen, A.S.; Arvidsson, H.S.; Munch, I.C.; Søndergaard, L.; Larsen, M. Retinal oxygen saturation in patients with systemic hypoxemia. Investig. Ophthalmol. Vis. Sci. 2011, 52, 5064–5067. [Google Scholar] [CrossRef][Green Version]
  72. Eliasdottir, T.S.; Bragason, D.; Hardarson, S.H.; Vacchiano, C.; Gislason, T.; Kristjansdottir, J.V.; Kristjansdottir, G.; Stefánsson, E. Retinal oximetry measures systemic hypoxia in central nervous system vessels in chronic obstructive pulmonary disease. PLoS ONE 2017, 12, e0174026. [Google Scholar] [CrossRef]
  73. Geirsdottir, A.; Palsson, O.; Hardarson, S.H.; Olafsdottir, O.B.; Kristjansdottir, J.V.; Stefánsson, E. Retinal vessel oxygen saturation in healthy individuals. Investig. Ophthalmol. Vis. Sci. 2012, 53, 5433–5442. [Google Scholar] [CrossRef][Green Version]
  74. Yang, W.; Fu, Y.; Dong, Y.; Lin, L.; Huang, X.; Li, Y.; Lin, X.; Gao, Q. Retinal vessel oxygen saturation in a healthy young Chinese population. Acta Ophthalmol. 2016, 94, 373–379. [Google Scholar] [CrossRef]
  75. Palkovits, S.; Lasta, M.; Boltz, A.; Schmidl, D.; Kaya, S.; Hammer, M.; Marzluf, B.; Popa-Cherecheanu, A.; Frantal, S.; Schmetterer, L.; et al. Measurement of retinal oxygen saturation in patients with chronic obstructive pulmonary disease. Investig. Ophthalmol. Vis. Sci. 2013, 54, 1008–1013. [Google Scholar] [CrossRef][Green Version]
  76. Told, R.; Boltz, A.; Schmetterer, L.; Garh€ofer, G.; Sacu, S.; Schmidt-Erfurth, U.; Pollreisz, A. Method comparison of two non-invasive dual-wavelength spectrophotometric retinal oximeters in healthy young subjects during normoxia. Acta Ophthalmol. 2018, 96, e614–e618. [Google Scholar] [CrossRef][Green Version]
  77. Eliasdottir, T.S. Retinal oximetry and systemic arterial oxygen levels. Acta Ophthalmol. 2018, 96, 1–44. [Google Scholar] [CrossRef] [PubMed][Green Version]
  78. Werkmeister, R.M.; Schmidl, D.; Aschinger, G.; Doblhoff-Dier, V.; Palkovits, S.; Wirth, M.; Garhöfer, G.; Linsenmeier, R.A.; Leitgeb, R.A.; Schmetterer, L. Retinal oxygen extraction in humans. Retinal oxygen extraction in humans. Sci. Rep. 2015, 5, 15763. [Google Scholar] [CrossRef] [PubMed][Green Version]
  79. Bata, A.M.; Fondi, K.; Szegedi, S.; Aschinger, G.C.; Hommer, A.; Schmidl, D.; Chua, J.; Werkmeister, R.M.; Garhöfer, G.; Schmetterer, L. Age-Related Decline of Retinal Oxygen Extraction in Healthy Subjects. Investig. Ophthalmol. Vis. Sci. 2019, 60, 3162–3169. [Google Scholar] [CrossRef] [PubMed][Green Version]
  80. Olafsdottir, O.B.; Eliasdottir, T.S.; Kristjansdottir, J.V.; Hardarson, S.H.; Stefánsson, E. Retinal Vessel Oxygen Saturation during 100% Oxygen Breathing in Healthy Individuals. PLoS ONE 2015, 10, e0128780. [Google Scholar] [CrossRef][Green Version]
  81. Olafsdottir, O.B.; Hardarson, S.H.; Gottfredsdottir, M.S.; Harris, A.; Stefánsson, E. Retinal oximetry in primary open-angle glaucoma. Investig. Ophthalmol. Vis. Sci. 2011, 52, 6409–6413. [Google Scholar] [CrossRef][Green Version]
  82. Hammer, M.; Heller, T.; Jentsch, S.; Dawczynski, J.; Schweitzer, D.; Peters, S.; Schmidtke, K.U.; Müller, U.A. Retinal vessel oxygen saturation under flicker light stimulation in patients with nonproliferative diabetic retinopathy. Investig. Ophthalmol. Vis. Sci. 2012, 53, 4063–4068. [Google Scholar] [CrossRef][Green Version]
  83. Vandewalle, E.; Pinto, L.A.; Olafsdottir, O.B.; De Clerck, E.; Stalmans, P.; Van Calster, J.; Zeyen, T.; Stefánsson, E.; Stalmans, I. Oximetry in Glaucoma: Correlation of Metabolic Change With Structural and Functional Damage. Acta Ophthalmol. 2014, 92, 105–110. [Google Scholar] [CrossRef]
  84. Battu, R.; Mohan, A.; Khanna, A.; Kumar, A.; Shetty, R. Retinal Oxygen Saturation in Retinitis Pigmentosa and Macular Dystrophies in Asian-Indian Eyes. Investig. Ophthalmol. Vis. Sci. 2015, 56, 2798–2802. [Google Scholar] [CrossRef][Green Version]
  85. Eliasdottir, T.S.; Bragason, D.; Hardarson, S.H.; Kristjansdottir, G.; Stefánsson, E. Venous Oxygen Saturation Is Reduced and Variable in Central Retinal Vein Occlusion. Graefes Arch. Clin. Exp. Ophthalmol. 2015, 253, 1653–1661. [Google Scholar] [CrossRef]
  86. Rilvén, S.; Torp, T.L.; Grauslund, J. Retinal oximetry in patients with ischaemic retinal diseases. Acta Ophthalmol. 2017, 95, 119–127. [Google Scholar] [CrossRef]
  87. London, A.; Benhar, I.; Schwartz, M. The retina as a window to the brain-from eye research to CNS disorders. Nat. Rev. Neurol. 2013, 9, 44–53. [Google Scholar] [CrossRef]
  88. DeBuc, D.C.; Somfai, G.M.; Koller, A. Retinal microvascular network alterations: Potential biomarkers of cerebrovascular and neural diseases. Am. J. Physiol. Heart Circ. Physiol. 2017, 312, H201–H212. [Google Scholar] [CrossRef] [PubMed]
  89. Arthur, E.; Somfai, G.M.; Kostic, M.; Oropesa, S.; Santiesteban, C.M.; DeBuc, D.C. Distinguishing cognitive impairment by using singularity spectrum and lacunarity analysis of the retinal vascular network. Neurophotonics 2019, 6, 041109. [Google Scholar] [CrossRef] [PubMed]
  90. Kur, J.; Newman, E.A.; Chan-Ling, T. Cellular and physiological mechanisms underlying blood flow regulation in the retina and choroid in health and disease. Prog. Retin Eye Res. 2012, 31, 377–406. [Google Scholar] [CrossRef] [PubMed][Green Version]
  91. Einarsdottir, A.B.; Olafsdottir, O.B.; Hjaltason, H.; Hardarson, S.H. Retinal oximetry is affected in multiple sclerosis. Acta Ophthalmol. 2018, 96, 528–530. [Google Scholar] [CrossRef] [PubMed][Green Version]
  92. Olafsdottir, O.B.; Saevarsdottir, H.S.; Hardarson, S.H.; Hannesdottir, K.H.; Traustadottir, V.D.; Karlsson, R.A.; Einarsdottir, A.B.; Jonsdottir, K.D.; Stefansson, E.; Snaedal, J. Retinal oxygen metabolism in patients with mild cognitive impairment. Alzheimers Dement. 2018, 10, 340–345. [Google Scholar] [CrossRef]
  93. Jöbsis, F.F. Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 1977, 198, 1264–1267. [Google Scholar] [CrossRef]
  94. Jöbsis-vanderVliet, F.F.; Jöbsis, P.D. Biochemical and physiological basis of medical near-infrared spectroscopy. J. Biomed. Opt. 1999, 4, 397–402. [Google Scholar] [CrossRef]
  95. Matcher, S.J.; Cope, M.; Delpy, D.T. Use of the water absorption spectrum to quantify tissue chromophore concentration changes in near-infrared spectroscopy. Phys. Med. Biol. 1994, 38, 177–196. [Google Scholar] [CrossRef]
  96. Matcher, S.J. Signal quantification and localization in tissue near-infrared spectroscopy. In Handbook of Optical Biomedical Diagnostics; Tuchin, V.V., Ed.; SPIE Press: Bellingham, WA, USA, 2002; Chapter 9. [Google Scholar]
  97. Delpy, D.T.; Cope, M. Quantification in tissue near-infrared spectroscopy. Phil. Trans. R. Soc. Lond. B 1997, 352, 649–659. [Google Scholar] [CrossRef][Green Version]
  98. Chance, B. Spectrophotometry of intracellular respiratory pigments. Science 1954, 120, 767–775. [Google Scholar] [CrossRef] [PubMed]
  99. Cope, M.; Delpy, D.T. System for long-term measurement of cerebral blood and tissue oxygenation on newborn infants by near infra-red transillumination. Med. Biol. Eng. Comput. 1988, 26, 289–294. [Google Scholar] [CrossRef]
  100. Bickler, P.E.; Feiner, J.R.; Rollins, M.D. Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers. Anesth. Analg. 2013, 117, 813–823. [Google Scholar] [CrossRef] [PubMed]
  101. Phan, P.; Highton, D.; Lai, J.; Smith, M.; Elwell, C.; Tachtsidis, I. Multi-channel multi-distance broadband near-infrared spectroscopy system to measure the spatial response of cellular oxygen metabolism and tissue oxygenation. Biomed. Opt. Express 2016, 7, 4424–4440. [Google Scholar] [CrossRef] [PubMed]
  102. De Blasi, R.A.; Ferrari, M.; Natali, A.; Conti, G.; Mega, A.; Gasparetto, A. Noninvasive measurement of forearm blood flow and oxygen consumption by near-infrared spectroscopy. J. Appl. Physiol. 1994, 76, 1388–1393. [Google Scholar] [CrossRef]
  103. Feldmann, A.; Schmitz, R.; Erlacher, D. Near-infrared spectroscopy-derived muscle oxygen saturation on a 0% to 100% scale: Reliability and validity of the Moxy Monitor. J. Biomed. Opt. 2019, 24, 115001. [Google Scholar] [CrossRef]
  104. Patterson, M.S.; Chance, B.; Wilson, B.C. Time resolved reflectance and transmittance for the non-invasive measurement of tissue optical properties. Appl. Opt. 1989, 28, 2331–2336. [Google Scholar] [CrossRef]
  105. Fantini, S.; Franceschini, M.A.; Gratton, E. Semi-infinite-geometry boundary problem for light migration in highly scattering media—A frequency-domain study in the diffusion-approximation. J. Opt. Soc. Am. B Opt. Phys. 1994, 11, 2128–2138. [Google Scholar] [CrossRef][Green Version]
  106. Wilson, B.C.; Sevick, E.M.; Patterson, M.S.; Chance, B. Time-dependent optical spectroscopy and imaging for biomedical applications. Proc. IEEE 1992, 80, 918–930. [Google Scholar] [CrossRef]
  107. Suzuki, S.; Takasaki, S.; Ozaki, T.; Kobayashi, Y. Tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc. SPIE 1999, 3597, 582–592. [Google Scholar]
  108. Liu, H.; Boas, D.A.; Zhang, Y.; Yodh, A.G.; Chance, B. Determination of optical properties and blood oxygenation in tissue using continuous NIR light. Phys. Med. Biol. 1995, 40, 1983–1993. [Google Scholar] [CrossRef] [PubMed][Green Version]
  109. Al-Rawi, P.G.; Smielewski, P.; Hobbiger, H.; Ghosh, S.; Kirkpatrick, P.J. Assessment of spatially resolved spectroscopy during cardiopulmonary bypass. J. Biomed. Opt. 1999, 4, 208–216. [Google Scholar] [CrossRef] [PubMed][Green Version]
  110. Kovacsova, Z.; Bale, G.; Mitra, S.; de Roever, I.; Meek, J.; Robertson, N.; Tachtsidis, I. Investigation of Confounding Factors in Measuring Tissue Saturation with NIRS Spatially Resolved Spectroscopy. In Oxygen Transport to Tissue XL; Advances in Experimental Medicine and Biology Series; Thews, O., LaManna, J., Harrison, D., Eds.; Springer: Cham, Switzerland, 2018; Volume 1072. [Google Scholar]
  111. Sorensen, L.C.; Greisen, G. Precision of measurement of cerebral tissue oxygenation index using near-infrared spectroscopy in preterm neonates. J. Biomed. Opt. 2006, 11, 054005. [Google Scholar] [CrossRef] [PubMed]
  112. Wolf, M.; Ferrari, M.; Quaresima, V. Progress of near-infrared spectroscopy and topography for brain and muscle clinical applications. J. Biomed. Opt. 2007, 12, 062104. [Google Scholar] [PubMed]
  113. Star, W.M.; Marijnissen, J.P.A.; Gemert, M.J.C. Light dosimetry in optical phantoms and in tissues: I. Multiple flux and transport theory. Phys. Med. Biol. 1988, 33, 437–454. [Google Scholar] [CrossRef]
  114. MacLeod, D.B.; Ikeda, K.; Vacchiano, C.; Lobbestael, A.; Wahr, J.A.; Shaw, A.D. Development and validation of a cerebral oximeter capable of absolute accuracy. J. Cardiothorac. Vasc. Anesth. 2012, 26, 1007–1014. [Google Scholar] [CrossRef]
  115. Redford, D.; Paidy, S.; Kashif, F. Absolute and trend accuracy of a new regional oximeter in healthy volunteers during controlled hypoxia. Anesth. Analg. 2014, 119, 1315–1319. [Google Scholar] [CrossRef][Green Version]
  116. Ikeda, K.; MacLeod, D.B.; Grocott, H.P.; Moretti, E.W.; Ames, W.; Vacchiano, C. The accuracy of a near-infrared spectroscopy cerebral oximetry device and its potential value for estimating jugular venous oxygen saturation. Anesth. Analg. 2014, 119, 1381–1392. [Google Scholar] [CrossRef][Green Version]
  117. Weigl, W.; Milej, D.; Janusek, D.; Wojtkiewicz, S.; Sawosz, P.; Kacprzak, M.; Gerega, A.; Maniewski, R.; Liebert, A. Application of optical methods in the monitoring of traumatic brain injury: A review. J. Cereb. Blood Flow Metab. 2016, 36, 1825–1843. [Google Scholar] [CrossRef]
  118. Mathieu, F.; Khellaf, A.; Ku, J.C.; Donnelly, J.; Thelin, E.P.; Zeiler, F.A. Continuous Near-Infrared Spectroscopy Monitoring in Adult Traumatic Brain Injury: A Systematic Review. J. Neurosurg. Anesthesiol. 2019, 10, 1097. [Google Scholar] [CrossRef]
  119. Davie, S.N.; Grocott, H.P. Impact of extracranial contamination on regional cerebral oxygen saturation: A comparison of three cerebral oximetry technologies. Anesthesiology 2012, 116, 834–840. [Google Scholar] [CrossRef] [PubMed][Green Version]
  120. Strangman, G.; Franceschini, M.A.; Boas, D.A. Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters. Neuroimage 2003, 18, 865–879. [Google Scholar] [CrossRef]
  121. Bozkurt, A.; Rosen, A.; Rosen, H.; Onaral, B. A portable near infrared Spectroscopy system for bedside monitoring of newborn brain. Biomed. Eng. Online 2005, 29, 29. [Google Scholar] [CrossRef] [PubMed][Green Version]
  122. Delpy, D.T.; Cope, M.; van der Zee, P.; Arridge, S.; Wray, S.; Wyatt, J. Estimation of optical pathlength through tissue from direct time of flight measurement. Phys. Med. Biol. 1988, 33, 1433–1442. [Google Scholar] [CrossRef][Green Version]
  123. Villringer, A.; Dirnagl, U. Coupling of brain activity and cerebral blood flow: Basis of functional neuroimaging. Cerebrovasc. Brain Metab. Rev. 1995, 7, 240–276. [Google Scholar] [PubMed]
  124. Mackert, B.M.; Wübbeler, G.; Leistner, S.; Uludag, K.; Obrig, H.; Villringer, A.; Trahms, L.; Curio, G. Neurovascular coupling analyzed non-invasively in the human brain. Neuroreport 2004, 15, 63–66. [Google Scholar] [CrossRef]
  125. Scholkmann, F.; Kleiser, S.; Metz, A.J.; Zimmermann, R.; Mata Pavia, J.; Wolf, U.; Wolf, M. A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology. Neuroimage 2014, 85, 6–27. [Google Scholar] [CrossRef]
  126. Lee, C.W.; Cooper, R.J.; Austin, T. Diffuse optical tomography to investigate the newborn brain. Pediatr. Res. 2017, 82, 376–386. [Google Scholar] [CrossRef][Green Version]
  127. Doulgerakis, M.; Eggebrecht, A.T.; Dehghani, H. High-density functional diffuse optical tomography based on frequency-domain measurements improves image quality and spatial resolution. Neurophotonics 2019, 6, 035007. [Google Scholar] [CrossRef]
  128. Liao, S.M.; Ferradal, S.L.; White, B.R.; Gregg, N.; Inder, T.E.; Culver, J.P. High-density diffuse optical tomography of term infant visual cortex in the nursery. J. Biomed. Opt. 2012, 17, 081414. [Google Scholar] [CrossRef][Green Version]
  129. Ferradal, S.L.; Liao, S.M.; Eggebrecht, A.T.; Shimony, J.S.; Inder, T.E.; Culver, J.P.; Smyser, C.D. Functional Imaging of the Developing Brain at the Bedside Using Diffuse Optical Tomography. Cereb. Cortex 2016, 26, 1558–1568. [Google Scholar] [CrossRef] [PubMed][Green Version]
  130. Karen, T.; Kleiser, S.; Ostojic, D.; Isler, H.; Guglielmini, S.; Bassler, D.; Wolf, M.; Scholkmann, F. Cerebral hemodynamic responses in preterm-born neonates to visual stimulation: Classification according to subgroups and analysis of frontotemporal–occipital functional connectivity. Neurophotonics 2019, 6, 045005. [Google Scholar] [CrossRef] [PubMed][Green Version]
  131. Wood, M.D.; Khan, J.; Lee, K.F.H.; Maslove, D.M.; Muscedere, J.; Hunt, M.; Scott, S.H.; Day, A.; Jacobson, J.; Ball, I.; et al. Assessing the relationship between near-infrared spectroscopy-derived regional cerebral oxygenation and neurological dysfunction in critically ill adults: A prospective observational multicentre protocol, on behalf of the Canadian Critical Care Trials Group. BMJ Open 2019, 25, e029189. [Google Scholar]
  132. Wood, M.D.; Jacobson, J.A.; Maslove, D.M.; Muscedere, J.G.; Boyd, J.G. Cerebral Oxygenation and Neurological Outcomes Following Critical Illness [CONFOCAL] Research Group. The physiological determinants of near-infrared spectroscopy-derived regional cerebral oxygenation in critically ill adults. Intensive Care Med. Exp. 2019, 7, 23. [Google Scholar] [CrossRef] [PubMed][Green Version]
  133. Bendahan, N.; Neal, O.; Ross-White, A.; Muscedere, J.; Boyd, J.G. Relationship Between Near-Infrared Spectroscopy-Derived Cerebral Oxygenation and Delirium in Critically Ill Patients: A Systematic Review. J. Intensive Care Med. 2019, 34, 514–520. [Google Scholar] [CrossRef] [PubMed]
  134. He, L.; Baker, W.B.; Milej, D.; Kavuri, V.C.; Mesquita, R.C.; Busch, D.R.; Abramson, K.; Jiang, J.Y.; Diop, M.; St. Lawrence, K.; et al. Noninvasive continuous optical monitoring of absolute cerebral blood flow in critically ill adults. Neurophotonics 2018, 5, 045006. [Google Scholar]
  135. La Cour, A.; Greisen, G.; Hyttel-Sorensen, S. In vivo validation of cerebral near-infrared spectroscopy: A review. Neurophotonics 2018, 5, 040901. [Google Scholar] [CrossRef]
  136. Rosenthal, G.; Furmanov, A.; Itshayek, E.; Shoshan, Y.; Singh, V. Assessment of a noninvasive cerebral oxygenation monitor in patients with severe traumatic brain injury: Clinical article. J. Neurosurg. 2014, 120, 901–907. [Google Scholar] [CrossRef]
  137. Ito, H.; Kanno, I.; Iida, H.; Hatazawa, J.; Shimosegawa, E.; Tamura, H.; Okudera, T. Arterial fraction of cerebral blood volume in humans measured by positron emission tomography. Ann. Nucl Med. 2001, 15, 111–116. [Google Scholar] [CrossRef]
  138. Kreeger, R.N.; Ramamoorthy, C.; Nicolson, S.C.; Ames, W.A.; Hirsch, R.; Peng, L.F.; Glatz, A.C.; Hill, K.D.; Hoffman, J.; Tomasson, J.; et al. Evaluation of pediatric near-infrared cerebral oximeter for cardiac disease. Ann. Thorac. Surg. 2012, 94, 1527–1533. [Google Scholar] [CrossRef]
  139. Kussman, B.D.; Laussen, P.C.; Benni, P.B.; McGowan, F.X., Jr.; McElhinney, D.B. Cerebral Oxygen Saturation in Children with Congenital Heart Disease and Chronic Hypoxemia. Anesth. Analg. 2017, 125, 234–240. [Google Scholar] [CrossRef] [PubMed]
  140. Benni, P.B.; MacLeod, D.; Ikeda, K.; Lin, H.M. A validation method for near-infrared spectroscopy based tissue oximeters for cerebral and somatic tissue oxygen saturation measurements. J. Clin. Monit. Comput. 2018, 32, 269–284. [Google Scholar] [CrossRef] [PubMed][Green Version]
  141. Leal-Noval, S.R.; Cayuela, A.; Arellano-Orden, V.; Marín-Caballos, A.; Padilla, V.; Ferrándiz-Millón, C.; Corcia, Y.; García-Alfaro, C.; Amaya-Villar, R.; Murillo-Cabezas, F. Invasive and noninvasive assessment of cerebral oxygenation in patients with severe traumatic brain injury. Intensive Care Med. 2010, 36, 1309–1317. [Google Scholar] [CrossRef] [PubMed]
  142. Esnault, P.; Boret, H.; Montcriol, A.; Carre, E.; Prunet, B.; Bordes, J.; Simon, P.; Joubert, C.; Dagain, A.; Kaiser, E.; et al. Assessment of cerebral oxygenation in neurocritical care patients: Comparison of a new four wavelengths forehead regional saturation in oxygen sensor [EQUANOX] with brain tissue oxygenation. A prospective observational study. Minerva Anestesiol. 2015, 81, 876–884. [Google Scholar] [PubMed]
  143. Arri, S.J.; Muehlemann, T.; Biallas, M.; Bucher, H.U.; Wolf, M. Precision of cerebral oxygenation and hemoglobin concentration measurements in neonates measured by near-infrared spectroscopy. J. Biomed. Opt. 2011, 16, 047005. [Google Scholar] [CrossRef][Green Version]
  144. Kleiser, S.; Ostojic, D.; Nasseri, N.; Isler, H.; Bucher, H.U.; Bassler, D.; Wolf, M.; Scholkmann, F.; Karen, T. In vivo precision assessment of a near-infrared spectroscopy-based tissue oximeter [OxyPrem v1.3] in neonates considering systemic hemodynamic fluctuations. J. Biomed. Opt. 2018, 23, 067003. [Google Scholar] [CrossRef][Green Version]
  145. Murkin, J.M.; Adams, S.J.; Novick, R.J.; Quantz, M.; Bainbridge, D.; Iglesias, I.; Cleland, A.; Schaefer, B.; Irwin, B.; Fox, S. Monitoring brain oxygen saturation during coronary bypass surgery: A randomized, prospective study. Anesth. Analg. 2007, 104, 51–58. [Google Scholar] [CrossRef]
  146. Slater, J.P.; Guarino, T.; Stack, J.; Vinod, K.; Bustami, R.T.; Brown, J.M., 3rd; Rodriguez, A.L.; Magovern, C.J.; Zaubler, T.; Freundlich, K.; et al. Cerebral oxygen desaturation predicts cognitive decline and longer hospital stay after cardiac surgery. Ann. Thorac. Surg. 2009, 87, 36–44. [Google Scholar] [CrossRef]
  147. Vohra, H.A.; Modi, A.; Ohri, S.K. Does use of intra-operative cerebral regional oxygen saturation monitoring during cardiac surgery lead to improved clinical outcomes? Interact. Cardiovasc. Thorac. Surg. 2009, 9, 318–322. [Google Scholar] [CrossRef][Green Version]
  148. Zheng, F.; Sheinberg, R.; Yee, M.S.; Ono, M.; Zheng, Y.; Hogue, C.W. Cerebral near-infrared spectroscopy monitoring and neurologic outcomes in adult cardiac surgery patients: A systematic review. Anesth. Analg. 2013, 116, 663–676. [Google Scholar] [CrossRef][Green Version]
  149. Holmgaard, F.; Vedel, A.G.; Rasmussen, L.S.; Paulson, O.B.; Nilsson, J.C.; Ravn, H.B. The association between postoperative cognitive dysfunction and cerebral oximetry during cardiac surgery: A secondary analysis of a randomised trial. Br. J. Anaesth. 2019, 123, 196–205. [Google Scholar] [CrossRef]
  150. Uysal, S.; Lin, H.M.; Trinh, M.; Park, C.H.; Reich, D.L. Optimizing cerebral oxygenation in cardiac surgery: A randomized controlled trial examining neurocognitive and perioperative outcomes. J. Thorac. Cardiovasc. Surg. 2020, 159, 943–953.e3. [Google Scholar] [CrossRef]
  151. Yu, Y.; Zhang, K.; Zhang, L.; Zong, H.; Meng, L.; Han, R. Cerebral near-infrared spectroscopy [NIRS] for perioperative monitoring of brain oxygenation in children and adults. Cochrane Database Syst. Rev. 2018, 1, CD010947. [Google Scholar] [CrossRef]
  152. Hyttel-Sorensen, S.; Pellicer, A.; Alderliesten, T.; Austin, T.; van Bel, F.; Benders, M.; Claris, O.; Dempsey, E.; Franz, A.R.; Fumagalli, M.; et al. Cerebral near infrared spectroscopy oximetry in extremely preterm infants: Phase II randomised clinical trial. BMJ 2015, 350, g7635. [Google Scholar] [CrossRef][Green Version]
  153. Plomgaard, A.M.; van Oeveren, W.; Petersen, T.H.; Alderliesten, T.; Austin, T.; van Bel, F.; Benders, M.; Claris, O.; Dempsey, E.; Franz, A.; et al. The SafeBoosC II randomized trial: Treatment guided by near-infrared spectroscopy reduces cerebral hypoxia without changing early biomarkers of brain injury. Pediatr. Res. 2016, 79, 528–535. [Google Scholar] [CrossRef] [PubMed]
  154. Plomgaard, A.M.; Alderliesten, T.; van Bel, F.; Benders, M.; Claris, O.; Cordeiro, M.; Dempsey, E.; Fumagalli, M.; Gluud, C.; Hyttel-Sorensen, S.; et al. No neurodevelopmental benefit of cerebral oximetry in the first randomised trial [SafeBoosC II] in preterm infants during the first days of life. Acta Paediatr. 2019, 108, 275–281. [Google Scholar] [CrossRef] [PubMed]
  155. Fischer, G.W.; Lin, H.M.; Krol, M.; Galati, M.F.; Di Luozzo, G.; Griepp, R.B.; Reich, D.L. Noninvasive cerebral oxygenation may predict outcome in patients undergoing aortic arch surgery. J. Thorac. Cardiovasc. Surg. 2011, 141, 815–821. [Google Scholar] [CrossRef] [PubMed][Green Version]
  156. Wen, S.; Peng, A.; Boyle, S.; Cai, S.; Pope, L.; Tran, M.T.; Short, T.G.; Aneman, A.; Jaeger, M.; Chuan, A. A pilot study using preoperative cerebral tissue oxygen saturation to stratify cardiovascular risk in major non-cardiac surgery. Anaesth. Intensive Care 2017, 45, 202–209. [Google Scholar] [CrossRef] [PubMed][Green Version]
  157. Pfister, D.; Siegemund, M.; Dell-Kuster, S.; Smielewski, P.; Rüegg, S.; Strebel, S.P.; Marsch, S.C.U.; Pargger, H.; Pargger, L.A. Cerebral perfusion in sepsis-associated delirium. Crit. Care 2008, 12, R63. [Google Scholar] [CrossRef] [PubMed][Green Version]
  158. Vaskó, A.; Siró, P.; László, I.; Szatmári, S.; Molnár, L.; Fülesdi, B.; Molnár, C. Assessment of cerebral tissue oxygen saturation in septic patients during acetazolamide provocation—A near infrared spectroscopy study. Acta Physiol. Hung. 2014, 101, 32–39. [Google Scholar] [CrossRef]
  159. Wood, M.; Song, A.; Maslove, D.; Ferri, C.; Howes, D.; Muscedere, J.; Boyd, J.G. Brain Tissue Oxygenation in Patients with Septic Shock: A Feasibility Study. Can. J. Neurol. Sci. 2016, 43, 65–73. [Google Scholar] [CrossRef]
  160. Wood, M.D.; Maslove, D.M.; Muscedere, J.G.; Day, A.G.; Gordon Boyd, J. Low brain tissue oxygenation contributes to the development of delirium in critically ill patients: A prospective observational study. J. Crit. Care. 2017, 41, 289–295. [Google Scholar] [CrossRef] [PubMed]
  161. Davies, D.J.; Su, Z.; Clancy, M.T.; Lucas, S.J.E.; Dehghani, H.; Logan, A.; Belli, A. Near-Infrared Spectroscopy in the Monitoring of Adult Traumatic Brain Injury: A Review. J. Neurotrauma 2015, 32, 933–941. [Google Scholar] [CrossRef] [PubMed]
  162. Panerai, R.B. Assessment of cerebral pressure autoregulation in humans—a review of measurement methods. Physiol. Meas. 1998, 19, 305–338. [Google Scholar] [CrossRef] [PubMed]
  163. Willie, C.K.; Tzeng, Y.-C.; Fisher, J.A.; Ainslie, P.N. Integrative regulation of human brain blood flow. J. Physiol. 2014, 592, 841–859. [Google Scholar] [CrossRef] [PubMed]
  164. Ainslie, P.N.; Brassard, P. Why is the neural control of cerebral autoregulation so controversial? F1000Prime Rep. 2014, 6, 14. [Google Scholar] [CrossRef]
  165. Liu, X.; Czosnyka, M.; Donnelly, J.; Budohoski, K.P.; Varsos, G.V.; Nasr, N.; Brady, K.M.; Reinhard, M.; Hutchinson, P.J.; Smielewski, P. Comparison of frequency and time domain methods of assessment of cerebral autoregulation in traumatic brain injury. J. Cereb. Blood Flow Metab. 2015, 35, 248–256. [Google Scholar] [CrossRef][Green Version]
  166. Steiner, L.A.; Pfister, D.; Strebel, S.P.; Radolovich, D.; Smielewski, P.; Czosnyka, M. Near-infrared spectroscopy can monitor dynamic cerebral autoregulation in adults. Neurocrit. Care 2009, 10, 122–128. [Google Scholar] [CrossRef][Green Version]
  167. Brady, K.; Joshi, B.; Zweifel, C.; Smielewski, P.; Czosnyka, M.; Easley, R.B.; Hogue, C.W., Jr. Real-time continuous monitoring of cerebral blood flow autoregulation using near-infrared spectroscopy in patients undergoing cardiopulmonary bypass. Stroke 2010, 41, 1951–1956. [Google Scholar] [CrossRef]
  168. Sen, A.N.; Gopinath, S.P.; Robertson, C.S. Clinical application of near-infrared spectroscopy in patients with traumatic brain injury: A review of the progress of the field. Neurophotonics 2016, 3, 031409. [Google Scholar] [CrossRef]
  169. Highton, D.; Ghosh, A.; Tachtsidis, I.; Panovska-Griffiths, J.; Elwell, C.E.; Smith, M. Monitoring cerebral autoregulation after brain injury: Multimodal assessment of cerebral slow-wave oscillations using near-infrared spectroscopy. Anesth. Analg. 2015, 121, 198–205. [Google Scholar] [CrossRef] [PubMed][Green Version]
  170. Liu, X.; Czosnyka, M.; Donnelly, J.; Cardim, D.; Cabeleira, M.; Lalou, D.A.; Hu, X.; Hutchinson, P.J.; Smielewski, P. Assessment of cerebral autoregulation indices—A modelling perspective. Sci. Rep. 2020, 10, 9600. [Google Scholar] [CrossRef] [PubMed]
  171. Brady, K.M.; Mytar, J.O.; Lee, J.K.; Cameron, D.E.; Vricella, L.A.; Thompson, W.R.; Hogue, C.W.; Easley, R.B. Monitoring cerebral blood flow pressure autoregulation in pediatric patients during cardiac surgery. Stroke 2010, 41, 1957–1962. [Google Scholar] [CrossRef] [PubMed][Green Version]
  172. Tichauer, K.M.; Elliott, J.T.; Hadway, J.A.; Lee, D.S.; Lee, T.Y.; St Lawrence, K. Using near-infrared spectroscopy to measure cerebral metabolic rate of oxygen under multiple levels of arterial oxygenation in piglets. J. Appl. Physiol. 2010, 109, 878–885. [Google Scholar] [CrossRef] [PubMed][Green Version]
  173. Kim, H.Y.; Seo, K.; Jeon, H.J.; Lee, U.; Lee, H. Application of Functional Near-Infrared Spectroscopy to the Study of Brain Function in Humans and Animal Models. Mol. Cells 2017, 40, 523–532. [Google Scholar] [CrossRef] [PubMed][Green Version]
  174. Nakamura, S.; Walker, D.W.; Wong, F.Y. Cerebral haemodynamic response to somatosensory stimulation in near-term fetal sheep. J. Physiol. 2017, 595, 1289–1303. [Google Scholar] [CrossRef][Green Version]
  175. El-Sabbagh, A.M.; Gray, B.W.; Shaffer, A.W.; Bryner, B.S.; Church, J.T.; McLeod, J.S.; Zakem, S.; Perkins, E.M.; Shellhaas, R.A.; Barks, J.D.E.; et al. Cerebral Oxygenation of Premature Lambs Supported by an Artificial Placenta. ASAIO J. 2018, 64, 552–556. [Google Scholar] [CrossRef]
  176. Kurita, T.; Kawashima, S.; Morita, K.; Nakajima, Y. Spinal cord autoregulation using near-infrared spectroscopy under normal, hypovolemic, and post-fluid resuscitation conditions in a swine model: A comparison with cerebral autoregulation. J. Intensive Care 2020, 8, 27. [Google Scholar] [CrossRef][Green Version]

Figure 1. Molar extinction coefficients of oxygenated [blue curve] and deoxygenated hemoglobin [red curve] as a function of the wavelength in the visible and near infrared regions. Prepared by Scott Prahl from a variety of sources [3] and is presented with permission from Dr. Prahl.

Figure 1. Molar extinction coefficients of oxygenated [blue curve] and deoxygenated hemoglobin [red curve] as a function of the wavelength in the visible and near infrared regions. Prepared by Scott Prahl from a variety of sources [3] and is presented with permission from Dr. Prahl.

Figure 2. Reflection [left] and transmission [right] fiber-optic intravenous blood oximeters. The emission and measuring optic fibers are drawn in grey, the catheter wall is drawn in black.

Figure 2. Reflection [left] and transmission [right] fiber-optic intravenous blood oximeters. The emission and measuring optic fibers are drawn in grey, the catheter wall is drawn in black.

Figure 3. The measured parameters of the PPG pulses. The transmitted light through the tissue decreases during systole and increases during diastole. AC is the difference between the maximal [ID] and minimal [IS] light transmission through the tissue; DC is the mean light transmission during the pulse. This Figure was published in Yossef Hay et al. Sensors 2018 [41].

Figure 3. The measured parameters of the PPG pulses. The transmitted light through the tissue decreases during systole and increases during diastole. AC is the difference between the maximal [ID] and minimal [IS] light transmission through the tissue; DC is the mean light transmission during the pulse. This Figure was published in Yossef Hay et al. Sensors 2018 [41].

Figure 4. Schematic cross-section of the retina [grey], which includes cross-sections of a retinal vessel [arteriole or venule] and retinal capillaries [red], and possible trajectories of light that illuminates the retina. The incident light [broad black rays] illuminates the vessel and its neighborhood. The light that strikes the vessel and the neighboring tissue is scattered and backscattered by the blood and the tissue [thin rays]. Most of the scattered/reflected light from the neighboring tissue is not affected by absorption in the blood, but some photons are affected by blood in the retinal vessel or the retinal capillaries.

Figure 4. Schematic cross-section of the retina [grey], which includes cross-sections of a retinal vessel [arteriole or venule] and retinal capillaries [red], and possible trajectories of light that illuminates the retina. The incident light [broad black rays] illuminates the vessel and its neighborhood. The light that strikes the vessel and the neighboring tissue is scattered and backscattered by the blood and the tissue [thin rays]. Most of the scattered/reflected light from the neighboring tissue is not affected by absorption in the blood, but some photons are affected by blood in the retinal vessel or the retinal capillaries.

Figure 5. Schematic presentation of the relationship between cerebral blood flow CBF and mean arterial pressure MAP, affected by autoregulation. The autoregulation is only effective between two MAP limits, presented by dashed lines. For MAP lower than the left limit, the brain tissue is ischemic, and for MAP greater than the right limit, the brain tissue is hyperemic, potentially leading to edema with its adverse consequences.

Figure 5. Schematic presentation of the relationship between cerebral blood flow CBF and mean arterial pressure MAP, affected by autoregulation. The autoregulation is only effective between two MAP limits, presented by dashed lines. For MAP lower than the left limit, the brain tissue is ischemic, and for MAP greater than the right limit, the brain tissue is hyperemic, potentially leading to edema with its adverse consequences.


© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution [CC BY] license [//creativecommons.org/licenses/by/4.0/].

Does fMRI detect blood oxygen levels?

Blood oxygen level–dependent [BOLD] functional MRI [fMRI] is the most commonly used method. It measures changes in blood flow by detecting changes in intravascular oxyhemoglobin concentration.

Which neuroimaging techniques is a measure of the deoxygenated blood in parts of the brain?

Blood-oxygenation-level-dependent functional magnetic resonance imaging is a noninvasive technique that has become increasingly popular in the neurosciences. It measures the proportion of oxygenated haemoglobin in specific areas of the brain, mirroring blood flow and therefore function.

What is the difference between EEG and fMRI?

As we have already noted, EEG signals are directly related to neuronal processing, whereas fMRI responses arise from subsequent changes in blood- oxygenation levels. There are other important differences as well that are rele- vant for combining the two types of data.

Which of the following does fMRI directly measure?

Functional magnetic resonance imaging [fMRI] measures the small changes in blood flow that occur with brain activity.